Adjustment of the evoked response sensitivity after hospital discharge in pacemaker patients with automatic ventricular threshold tracking activated

被引:6
作者
Schuchert, A [1 ]
Ventura, R [1 ]
Meinertz, T [1 ]
机构
[1] Univ Hamburg, Hosp Eppendorf, Med Clin, Dept Cardiol, D-20246 Hamburg, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2001年 / 24卷 / 02期
关键词
evoked response signal; lead polarization; automatic capture verification function Autocapture; VVIR pacemaker; pacemaker follow-up;
D O I
10.1046/j.1460-9592.2001.00212.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Automatic threshold tracking in cardiac pacemakers allows ventricular capture verification and self-adaptive pacing output regulation. The Autocapture algorithm detects the evoked response (ER) signal immediately after the pacing pulse to verify the efficacy of ventricular pacing. Before hospital delivery, the ER sensitivity must be programmed individually so that the pacemaker detects the ER signal adequately without sensing lead polarization. The aims of the study were to assess the frequency of patients in It hom Autocapture could be activated and whether the ER sensitivity had to be adjusted after hospital discharge. The study included 44 patients who received the VVIR pacemaker Regency SR+ (St. Jude Medical) connected to the model 1450 T pacing lead. ER signal, lead polarization, and ER sensitivity were evaluated before hospital discharge and 1, 3, and 6 months after implantation. The system recommended activating Autocapture in 42 of 44 patients. The mean ER signal was 8.4 +/- 1.2 mV at discharge, 9.0 +/- 3.9 mV at month 1, 8.9 +/- 4.9 mV at month 3, and 9.3 +/- 4.5 mV at month 6. Polarization was 1.0 +/- 0.1 mV at discharge, 1.1 +/- 0.5 mV at month 1, 1.1 +/- 0.2 mV at month 3, and 1.1 +/- 0.5 mV at month 6. Mean ER sensitivity was 3.7 +/- 1.8 mV at discharge, 4.0 +/- 1.8 mV after 1, 4.1 +/- 2.2 mV after 3, and 4.1 +/- 1.8 mV after 6 months. ER sensitivity could remain unadjusted in 14 patients. Programming to a less sensitivity ER setting from 2.9 +/- 1.2 mV to 4.3 +/- 1.5 mV was possible in 21 patients. Programming to a more sensitive ER setting from 4.1 +/- 1.1 mV to 2.5 +/- 0.9 mV was required in nine patients because of the decrease of the ER signal. The automatic threshold tracking algorithm Autocapture could be activated in 95% of patients. Programming to more sensitive ER settings was recommended in 21% of the patients after hospital discharge. Therefore, ER signal and polarization must be checked at each follow-up, as a decrease in ER signal amplitude can make reprogramming of the ER sensitivity necessary. There is no risk for the patient if the ER is not sensed, as high voltage backup stimulation is present.
引用
收藏
页码:212 / 216
页数:5
相关论文
共 15 条
[1]   THE VENTRICULAR DEPOLARIZATION GRADIENT - EFFECTS OF EXERCISE, PACING RATE, EPINEPHRINE, AND INTRINSIC HEART-RATE CONTROL ON THE RIGHT VENTRICULAR EVOKED-RESPONSE [J].
CALLAGHAN, F ;
VOLLMANN, W ;
LIVINGSTON, A ;
BOVEJA, B ;
ABELS, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (07) :1115-1130
[2]  
CERCAS A, 1997, PACE, V20, P1189
[3]   Automatic adjustment of pacemaker stimulation output correlated with continuously monitored capture thresholds:: A multicenter study [J].
Clarke, M ;
Liu, B ;
Schüller, H ;
Binner, L ;
Kennergren, C ;
Guerola, M ;
Weinmann, P ;
Ohm, OJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (08) :1567-1575
[4]   Pacing threshold trends and variability in modern tined leads assessed using high resolution automatic measurements: Conversion of pulse width into voltage thresholds [J].
Danilovic, D ;
Ohm, OJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (04) :567-587
[5]  
GREENBERG S, 1997, PACE, V20, P1172
[6]  
Lacroix D, 1998, PACE, V21, P1680, DOI 10.1111/j.1540-8159.1998.tb00262.x
[7]   Is this the natural history of the retention wire? A case report [J].
Lau, C ;
Nishimura, SC ;
Oxorn, D ;
Goldman, BS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (05) :1373-1376
[8]   The development of sinoatrial dysfunction in pacemaker patients with isolated atrioventricular block [J].
Morsi, A ;
Lau, C ;
Nishimura, S ;
Goldman, BS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (07) :1430-1434
[9]   CHANGES IN MYOCARDIAL THRESHOLD . PHYSIOLOGIC AND PHARMACOLOGIC FACTORS IN PATIENTS WITH IMPLANTED PACEMAKERS [J].
PRESTON, TA ;
FLETCHER, RD ;
LUCCHESI, BR ;
JUDGE, RD .
AMERICAN HEART JOURNAL, 1967, 74 (02) :235-&
[10]  
RIDDELL F, 1998, PACE, V21, P955