Retrofitted confocal laser scanner for a commercial inverted fluorescence microscope

被引:13
作者
Bigelow, CE
Harkrider, CJ
Conover, DL
Foster, TH
Georgakoudi, I
Mitra, S
Nichols, MG
Rajadhyaksha, M
机构
[1] Univ Rochester, Inst Opt, Rochester, NY 14627 USA
[2] Univ Rochester, Dept Radiol, Rochester, NY 14642 USA
[3] Univ Rochester, Inst Opt, Dept Radiol, Rochester, NY 14642 USA
[4] Univ Rochester, Dept Phys & Astron, Rochester, NY 14642 USA
[5] Univ Rochester, Dept Biochem & Biophys, Rochester, NY 14642 USA
[6] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Wellman Labs Photomed, Boston, MA 02114 USA
关键词
D O I
10.1063/1.1382637
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe the design and implementation of an inverted laser scanning confocal fluorescence microscope utilizing the commercial Nikon Diaphot TMD platform. An external confocal scanner was retrofitted through the video side port of the Diaphot. With 10x, 0.5 NA dry and 60x, 1.4 NA oil immersion objectives, the depth discrimination is 5.8 mum and 0.8 mum, respectively, as determined by derivatives of fluorescence edge responses measured in liquid samples of rhodamine 6G dissolved in DMSO. We present sample edge response curves and representative confocal fluorescence images of tumor cells in monolayer culture. (C) 2001 American Institute of Physics.
引用
收藏
页码:3407 / 3410
页数:4
相关论文
共 7 条
[1]  
[Anonymous], 1990, CONFOCAL MICROSCOPY
[2]  
Brenner M., COMMUNICATION
[3]   ABERRATIONS IN CONFOCAL FLUORESCENCE MICROSCOPY INDUCED BY MISMATCHES IN REFRACTIVE-INDEX [J].
HELL, S ;
REINER, G ;
CREMER, C ;
STELZER, EHK .
JOURNAL OF MICROSCOPY, 1993, 169 :391-405
[4]  
PELSUE K, 1983, P SOC PHOTO-OPT INS, V370, P70
[5]  
Vetterling W. T, 2002, NUMERICAL RECIPES C
[6]  
Webb RH, 1999, METHOD ENZYMOL, V307, P3
[7]  
WEBB RH, 1990, HDB BIOL CONFOCAL MI