Detection of simulated lung nodules with computed radiography: Effects of nodule size, local optical density, global object thickness, and exposure

被引:7
作者
KimmeSmith, C
Hart, EM
Goldin, JG
Johnson, TD
Terwilliger, R
Aberle, DR
机构
[1] Department of Radiological Sciences, University of California, Los Angeles, School of Medicine, Los Angeles, CA
[2] Department of Radiological Sciences, University of California, Los Angeles, School of Medicine, Los Angeles, CA 90095-1721
关键词
computed radiography; image quality; lung nodule; exposure;
D O I
10.1016/S1076-6332(96)80412-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. We quantified differences in the detection of simulated lung nodules on computed radiographs on the basis of variations in nodule size, local contrast, body habitus (global contrast), and exposure. Methods. A step-wedge phantom was developed to simulate the attenuation ranges of the lung, retrocardiac, and subdiaphragmatic regions of the adult human chest. Additional Lucite wedges were used to simulate two different body thicknesses and to provide variable structural noise. Soft-tissue-equivalent nodules of 3-mm and 5-mm diameter that resulted in 10% differences in attenuation from lung equivalence were embedded in lung-equivalent material. By superimposing the sheets in various positions, 84 unique nodule configurations containing eight nodules per image were exposed on a computed radiography system. Computed radiographs were acquired at two different exposures approximating standard exposure and underexposure. For each resulting phantom image, seven observers scored the presence or absence of a nodule within individual cells of a 5 x 5 grid matrix. Results. True-positive fractions for 3-mm-diameter nodules were very low across all conditions. True-positive fractions for 5-mm-diameter nodules varied from 0.23 to 0.98. Significant differences in the conspicuity of 5-mm nodules depended on differences in phantom thickness and differences in the locations of nodules within lung-, retrocardiac-, or subdiaphragmatic-equivalent regions. Accuracy in detecting nodules was significantly lower at lower exposures when nodules were located in the subdiaphragmatic-equivalent region. Conclusion. On computed radiographs, small nodules (5-mm diameter) can be reliably detected when they are located in areas of high or moderate surrounding local contrast, such as the lung or mediastinal regions. Detection of nodules decreases in regions of lower optical density corresponding to the subdiaphragmatic regions of the chest. The decrease in nodule detectability is greatest under conditions that simulate large body thickness and underexposure.
引用
收藏
页码:735 / 741
页数:7
相关论文
共 15 条
[1]   MISSED BRONCHOGENIC-CARCINOMA - RADIOGRAPHIC FINDINGS IN 27 PATIENTS WITH A POTENTIALLY RESECTABLE LESION EVIDENT IN RETROSPECT [J].
AUSTIN, JHM ;
ROMNEY, BM ;
GOLDSMITH, LS .
RADIOLOGY, 1992, 182 (01) :115-122
[2]   COMPARISON OF RECEIVER OPERATING CHARACTERISTIC AND FORCED-CHOICE OBSERVER PERFORMANCE-MEASUREMENT METHODS [J].
BURGESS, AE .
MEDICAL PHYSICS, 1995, 22 (05) :643-655
[3]   DIGITAL AND CONVENTIONAL CHEST IMAGING - A MODIFIED ROC STUDY OF OBSERVER PERFORMANCE USING SIMULATED NODULES [J].
CHAKRABORTY, DP ;
BREATNACH, ES ;
YESTER, MV ;
SOTO, B ;
BARNES, GT ;
FRASER, RG .
RADIOLOGY, 1986, 158 (01) :35-39
[4]   DIGITAL RADIOGRAPHY WITH PHOTOSTIMULABLE STORAGE PHOSPHORS - CONTROL OF DETECTOR LATITUDE IN CHEST IMAGING [J].
CHOTAS, HG ;
RAVIN, CE .
INVESTIGATIVE RADIOLOGY, 1992, 27 (10) :822-828
[5]  
DIXON WJ, 1983, INTRO STATISTICAL AN
[6]   THRESHOLD PERCEPTION PERFORMANCE WITH COMPUTED AND SCREEN-FILM RADIOGRAPHY - IMPLICATIONS FOR CHEST RADIOGRAPHY [J].
DOBBINS, JT ;
RICE, JJ ;
BEAM, CA ;
RAVIN, CE .
RADIOLOGY, 1992, 183 (01) :179-187
[7]  
HEDGE LV, 1985, STAT METHODS META AN
[8]   EFFECTS OF REDUCED EXPOSURE ON COMPUTED RADIOGRAPHY - COMPARISON OF NODULE DETECTION ACCURACY WITH CONVENTIONAL AND ASYMMETRIC SCREEN-FILM RADIOGRAPHS OF A CHEST PHANTOM [J].
KIMMESMITH, C ;
ABERLE, DR ;
SAYRE, JW ;
HART, EM ;
GREAVES, SM ;
BROWN, K ;
YOUNG, DA ;
DESERAN, MD ;
JOHNSON, T ;
JOHNSON, SL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (02) :269-273
[9]   LESION CONSPICUITY, STRUCTURED NOISE, AND FILM READER ERROR [J].
KUNDEL, HL ;
REVESZ, G .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1976, 126 (06) :1233-1238
[10]  
LEE KR, 1991, RADIOGRAPHICS, V11, P1013, DOI 10.1148/radiographics.11.6.1749846