Imaging performance of single-element CdZnTe detectors for digital radiography

被引:4
作者
Bertolucci, E [1 ]
Chirco, P
Conti, M
Marcello, L
Rossi, M
Russo, P
机构
[1] Univ Naples Federico II, Dipartimento Sci Fis, Naples, Italy
[2] Ist Nazl Fis Nucl, I-80125 Naples, Italy
[3] Univ Bologna, Dipartmento Fis, Bologna, Italy
[4] Ist Nazl Fis Nucl, I-40126 Bologna, Italy
关键词
D O I
10.1109/23.682417
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We measured the charge collection efficiency and energy resolution at 60 keV and 122 keV for single-element Cd0.8Zn0.2Te semiconductor detectors 2x10x1O mm(3) irradiated transverse to the electric field, in view of their possible use for digital radiography. Charge collection efficiency, energy resolution and charge collection time have been measured as a function of the bias voltage. The detector's imaging performance was evaluated, using a scanning procedure, by measurements (at 60 keV) of the line spread function and edge response function of the imaging system; metallic test objects and bone tissue were imaged. Image contrast values as low as C=10% were obtained for a minimum 5:1 S.N.R..
引用
收藏
页码:406 / 412
页数:7
相关论文
共 15 条
[1]   A RADIATION DETECTION SYSTEM FOR HIGH-ENERGY COMPUTERIZED-TOMOGRAPHY USING CDZNTE DETECTORS [J].
BALDAZZI, G ;
ROSSI, M ;
QUERZOLA, E ;
GUIDI, G ;
CHIRCO, P ;
SCANNAVINI, MG ;
ZANARINI, M ;
CASALI, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :575-579
[2]   Progress in developing focal-plane-multiplexer readout for large CdZnTe arrays for nuclear medicine applications [J].
Barber, HB ;
Marks, DG ;
Apotovsky, BA ;
Augustine, FL ;
Barrett, HH ;
Butler, JF ;
Dereniak, EL ;
Doty, FP ;
Eskin, JD ;
Hamilton, WJ ;
Matherson, KJ ;
Venzon, JE ;
Woolfenden, JM ;
Young, ET .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :262-265
[3]   CHARGE-TRANSPORT IN ARRAYS OF SEMICONDUCTOR GAMMA-RAY DETECTORS [J].
BARRETT, HH ;
ESKIN, JD ;
BARBER, HB .
PHYSICAL REVIEW LETTERS, 1995, 75 (01) :156-159
[4]   Timing properties of CdZnTe detectors for positron emission tomography [J].
Bertolucci, E ;
Conti, M ;
Curto, CA ;
Russo, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 400 (01) :107-112
[5]  
BUTLER JF, 1993, MATER RES SOC SYMP P, V302, P497, DOI 10.1557/PROC-302-497
[6]  
BUTLER JF, 1991, IEEE NUCL SCI S MED, P129
[7]  
HE Z, 1996, IEEE NUCL SCI S AN N, V1, P331
[8]   UNIPOLAR CHARGE SENSING WITH COPLANAR ELECTRODES - APPLICATION TO SEMICONDUCTOR-DETECTORS [J].
LUKE, PN .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :207-213
[9]   Performance of CdZnTe coplanar-grid gamma-ray detectors [J].
Luke, PN ;
Eissler, EE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1481-1486
[10]   The use of pulse processing techniques to improve the performance of Cd1-xZnxTe gamma-ray spectrometers [J].
Lund, JC ;
Olsen, R ;
VanScyoc, JM ;
James, RB .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1411-1416