Electrode configuration and energy resolution in gamma-ray detectors

被引:53
作者
Luke, PN
机构
[1] Lawrence Berkeley Natl. Laboratory, Berkeley
关键词
D O I
10.1016/S0168-9002(96)00353-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Materials used for room-temperature gamma-ray detectors exhibit substantial charge-collection deficiencies. As a result, the geometrical arrangement of charge-sensing electrodes on a detector can have a very large influence on its spectral performance. A discussion of the effects of electrode geometry on charge signal generation and energy resolution is presented. Various electrode configurations including spherical, coaxial, pixel, strip and coplanar-grid are examined in theoretical terms to evaluate their effectiveness and limitations in enhancing the performance of detectors with charge trapping.
引用
收藏
页码:232 / 237
页数:6