Correction for non-spectroscopic matrix effects in inductively coupled plasma-atomic emission spectroscopy by internal standardization using spectral lines of the same analyte

被引:36
作者
Al-Ammar, AS [1 ]
Barnes, RM [1 ]
机构
[1] Univ Massachusetts, Dept Chem, Lederle Grad Res Ctr, Amherst, MA 01003 USA
关键词
matrix effect correction; inductively coupled plasma-atomic emission spectroscopy; common analyte internal standardization (CAIS);
D O I
10.1016/S0584-8547(98)00190-6
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The common analyte internal standardization (CAIS) technique was extended to correct for non-spectroscopic matrix effects in inductively coupled plasma-atomic emission spectroscopy (ICP-AES) measurements. The technique is based on simultaneous measurement of two different spectral lines of the same analyte. A matrix correction factor is then estimated from its linear correlation with the ratio of intensities of these two measured lines. Experimental tests with four elements (Ba, La, Mg, and Mn) in three matrices (NaCl, H2SO4, HNO3) demonstrate a significant decrease (from 3 to 22 times) of the matrix effect after correction. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1583 / 1593
页数:11
相关论文
共 13 条
[1]  
Al-Ammar AS, 1998, ATOM SPECTROSC, V19, P18
[2]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[3]  
BROWNER RF, 1987, INDUCTIVELY COUPLE 2, P244
[4]   The determination of chromium, copper and nickel in groundwater using axial plasma inductively coupled plasma atomic emission spectrometry and proportional correction matrix effect reduction [J].
de Boer, JLM ;
van Leeuwen, W ;
Kohlmeyer, U ;
Breugem, PM .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1998, 360 (02) :213-218
[5]   Empirical procedure for the reduction of mixed-matrix effects in inductively coupled plasma atomic emission spectrometry using an internal standard and proportional correction [J].
deBoer, JLM ;
Velterop, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1996, 356 (06) :362-370
[6]   INTRA-ALKALI MATRIX EFFECTS IN THE INDUCTIVELY COUPLED PLASMA [J].
FAIRES, LM ;
APEL, CT ;
NIEMCZYK, TM .
APPLIED SPECTROSCOPY, 1983, 37 (06) :558-563
[7]   INFLUENCE OF OPERATING-CONDITIONS ON THE EFFECTS OF ACIDS IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
FERNANDEZ, A ;
MURILLO, M ;
CARRION, N ;
MERMET, JM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :217-221
[8]  
HELLINGESS D, 1994, ATOM SPECTROSC, V15, P254
[9]  
MAESSEN F, 1982, ICP INFORM NEWSL, V8, P21
[10]  
Paama L, 1997, ATOM SPECTROSC, V18, P119