Measurements of magnetism-related forces and torque moments affecting medical instruments, implants, and foreign objects during magnetic resonance imaging at all degrees of freedom

被引:11
作者
Planert, J
Modler, H
Vosshenrich, R
机构
[1] UNIV GOTTINGEN,INST MED PHYS & BIOPHYS,GOTTINGEN,GERMANY
[2] UNIV GOTTINGEN,DEPT RADIOG DIAGNOST 2,GOTTINGEN,GERMANY
关键词
force/torque moment sensor; MRI; magnetic field related forces; patient risks;
D O I
10.1118/1.597853
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We designed a novel force/torque moment transducer with six degrees of freedom. This new sensor is used for measuring the forces and torque moments exerted by static magnetic resonance imaging (MRI) fields on various medical implants, instruments, and ballistic shell fragments. To determine the location of strongest effect the inhomogeneity of the magnetic field between the entrance of the MRI-tube (portal), and the imaging section was measured. Force and torque moments of a sample of 45 objects were measured at two positions, at the portal, and in the center of the imaging section. We classified the objects regarding to the magnetism-related forces measured in the homogeneous part of the MRI tube. About one-third of the objects was non- or weakly magnetic, one-third moderately, and the last third was strongly magnetic. Forces within the inhomogeneous area of the magnetic field at the entrance part of the tube were up to half a magnitude larger than forces inside the homogeneous imaging section. Here the greatest potential risks to patients might occur. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:851 / 856
页数:6
相关论文
共 8 条
[1]   IMPORTANT CONSIDERATIONS IN MEASUREMENT OF ATTRACTIVE FORCE ON METALLIC IMPLANTS IN MR IMAGERS [J].
KAGETSU, NJ ;
LITT, AW .
RADIOLOGY, 1991, 179 (02) :505-508
[2]  
KELSEY CA, 1991, RADIOLOGY, V180, P280
[3]   POTENTIAL HAZARDS AND ARTIFACTS OF FERROMAGNETIC AND NONFERROMAGNETIC SURGICAL AND DENTAL MATERIALS AND DEVICES IN NUCLEAR MAGNETIC-RESONANCE IMAGING [J].
NEW, PFJ ;
ROSEN, BR ;
BRADY, TJ ;
BUONANNO, FS ;
KISTLER, JP ;
BURT, CT ;
HINSHAW, WS ;
NEWHOUSE, JH ;
POHOST, GM ;
TAVERAS, JM .
RADIOLOGY, 1983, 147 (01) :139-148
[4]   A MINIATURIZED FORCE TORQUE SENSOR WITH 6-DEGREES OF FREEDOM FOR DENTAL MEASUREMENTS [J].
PLANERT, J ;
MODLER, H ;
LUDECKE, K ;
EGER, M .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 (03) :241-248
[5]   MR IMAGING OF METALLIC IMPLANTS AND MATERIALS - A COMPILATION OF THE LITERATURE [J].
SHELLOCK, FG .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (04) :811-814
[6]   MR PROCEDURES AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - 1993 UPDATE [J].
SHELLOCK, FG ;
MORISOLI, S ;
KANAL, E .
RADIOLOGY, 1993, 189 (02) :587-599
[7]  
SHELLOCK FG, 1991, AM J NEURORADIOL, V12, P279
[8]   METALLIC BALLISTIC FRAGMENTS - MR IMAGING SAFETY AND ARTIFACTS [J].
TEITELBAUM, GP ;
YEE, CA ;
VANHORN, DD ;
KIM, HS ;
COLLETTI, PM .
RADIOLOGY, 1990, 175 (03) :855-859