Application of standardisation methods to correct the spectral differences induced by a fibre optic probe used for the near-infrared analysis of pharmaceutical tablets

被引:27
作者
Bouveresse, E
Casolino, C
de la Pezuela, C
机构
[1] Dipartimento Analisi & Tecnol Farmaceut Alimentar, I-16147 Genoa, Italy
[2] Univ Autonoma Barcelona, Dept Quim, Unidad Quim Analit, E-08193 Barcelona, Spain
关键词
near-infrared spectroscopy; instrument standardisation; calibration transfer; multivariate calibration; fibre optic module;
D O I
10.1016/S0731-7085(98)00169-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Near infrared spectroscopy has become very popular in the pharmaceutical industry because of many important practical advantages. With the help of powerful chemometric techniques, multivariate calibration models are developed, relating near-infrared spectra to the values to be modelled. However, because of small instrumental differences between near-infrared spectrometers, a calibration model can only be used with spectra collected on the same instrument, which represents a serious limitation for the use of near-infrared spectroscopy in the pharmaceutical industry. To deal with this important problem, a certain number of different standardisation approaches were proposed in the literature. In this article, an application of instrument standardisation methods is presented, where two different data measurement modules (internal measurement cell and external fibre optic module) of a near-infrared spectrometer must be standardised for the quantitative determination of an active compound in pharmaceutical tablets. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:35 / 42
页数:8
相关论文
共 32 条
[1]   IDENTIFICATION OF TABLET FORMULATIONS INSIDE BLISTER PACKAGES BY NEAR-INFRARED SPECTROSCOPY [J].
ALDRIDGE, PK ;
MUSHINSKY, RF ;
ANDINO, MM ;
EVANS, CL .
APPLIED SPECTROSCOPY, 1994, 48 (10) :1272-1276
[2]  
[Anonymous], 1993, J NEAR INFRARED SPEC, DOI DOI 10.1255/jnirs.25
[3]   CONTROL ANALYSIS OF A PHARMACEUTICAL PREPARATION BY NEAR-INFRARED REFLECTANCE SPECTROSCOPY - A COMPARATIVE-STUDY OF A SPINNING MODULE AND FIBER OPTIC PROBE [J].
BLANCO, M ;
COELLO, J ;
ITURRIAGA, H ;
MASPOCH, S ;
DELAPEZUELA, C ;
RUSSO, E .
ANALYTICA CHIMICA ACTA, 1994, 298 (02) :183-191
[4]   Standardisation of near-infrared spectrometric instruments: A review [J].
Bouveresse, E ;
Massart, DL .
VIBRATIONAL SPECTROSCOPY, 1996, 11 (01) :3-15
[5]   CALIBRATION TRANSFER ACROSS NEAR-INFRARED SPECTROMETRIC INSTRUMENTS USING SHENKS ALGORITHM - EFFECTS OF DIFFERENT STANDARDIZATION SAMPLES [J].
BOUVERESSE, E ;
MASSART, DL ;
DARDENNE, P .
ANALYTICA CHIMICA ACTA, 1994, 297 (03) :405-416
[6]   Detection, interpretation and correction of changes in the instrumental response of near-infrared monochromator instruments over time [J].
Bouveresse, E ;
Rutan, SC ;
VanderHeyden, Y ;
Penninckx, W ;
Massart, DL .
ANALYTICA CHIMICA ACTA, 1997, 348 (1-3) :283-301
[7]   MODIFIED ALGORITHM FOR STANDARDIZATION OF NEAR-INFRARED SPECTROMETRIC INSTRUMENTS [J].
BOUVERESSE, E ;
MASSART, DL ;
DARDENNE, P .
ANALYTICAL CHEMISTRY, 1995, 67 (08) :1381-1389
[8]   Improvement of the piecewise direct standardisation procedure for the transfer of NIR spectra for multivariate calibration [J].
Bouveresse, E ;
Massart, DL .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1996, 32 (02) :201-213
[9]  
BOUVERESSE E, 1996, ANALUSIS, V68, P981
[10]  
BOUVERESSE E, 1997, THESIS VRIJE U BRUSS