High sensitivity gas permeability measurement system for thin plastic films

被引:22
作者
Ranade, A
D'Souza, NA
Wallace, RM
Gnade, BE [1 ]
机构
[1] Univ Texas, Dept Elect Engn, Erik Jonsson Sch Engn & Comp Sci, Richardson, TX 75080 USA
[2] Univ N Texas, Dept Mat Sci & Engn, Denton, TX 76203 USA
基金
美国国家航空航天局;
关键词
D O I
10.1063/1.1823792
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have developed a system to quantitatively measure the permeation of gases through thin flexible substrates with high sensitivity. The system consists of two chambers, a high pressure side and an ultrahigh vacuum (UHV) side, separated by the flexible sample to be analyzed. The system is calibrated using a combination of a National Institute of Standards and Technology traceable calibrated He leak and a variable aperture calibrated orifice. The base total pressure for the UHV side is 1-3x10(-10) Torr. The partial pressure of individual gases that we are studying is <10(-10) Torr. The sample to be measured is secured between the two chambers using two 2.75 conflat flanges, two copper gaskets, and two indium "O" rings. The key factors that impact sensitivity and quantification are (1) reducing the residual partial pressure of the gas of interest to as low a value as possible on the UHV side of the system, (2) sealing the plastic sample between the two chambers with no detectable gas leakage around the sample or from the outside, (3) supporting the plastic substrate so that it can withstand 700 Torr pressure on the high pressure side, and (4) developing a calibration procedure that closely mimics the actual permeability measurement. The system allows us to measure permeation rates as low as 1x10(-6) g/m(2)-day for He, 1x10(-6) g/m(2)-day for O-2, and 5x10(-7) g/m(2)-day for Ar. (C) 2005 American Institute of Physics.
引用
收藏
页码:013902 / 1
页数:5
相关论文
共 14 条
[1]  
Atkins PW, 2001, PHYS CHEM
[2]   TESTING OF TURBOMOLECULAR PUMPS [J].
BASFORD, JA .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1992, 10 (04) :2623-2628
[3]   Gas permeation and lifetime tests on polymer-based barrier coatings [J].
Burrows, PE ;
Graff, GL ;
Gross, ME ;
Martin, PM ;
Hall, M ;
Mast, E ;
Bonham, C ;
Bennett, W ;
Michalski, L ;
Weaver, M ;
Brown, JJ ;
Fogarty, D ;
Sapochak, LS .
ORGANIC LIGHT-EMITTING MATERIALS AND DEVICES IV, 2000, 4105 :75-83
[4]  
Crank J., 1968, DIFFUSION POLYM
[5]  
DHOOT S, 2002, ENCY POLYM SCI TECHN
[6]   OBSERVATIONS ON BAYARD-ALPERT ION GAUGE SENSITIVITIES TO VARIOUS GASES [J].
FLAIM, TA ;
OWNBY, PD .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1971, 8 (05) :661-&
[7]   A LASER DRILLED APERTURE FOR USE IN AN ULTRAHIGH-VACUUM GAS DOSER [J].
HAGANS, PL ;
DEKOVEN, BM ;
WOMACK, JL .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1989, 7 (06) :3375-3377
[8]  
HERNANDEZ RJ, 2000, PLASTIC PACKAGING PR, P313
[9]   Measurement of the permeability of thin films [J].
Ianno, NJ ;
Makovicka, TJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (04) :2072-2073
[10]  
Nisato G., 2001, P 21 ANN AS DISPL 8, P1435