Development and validation of implantable sensors for monitoring function of prosthetic heart valves:: in vitro studies

被引:13
作者
Lanning, C
Shandas, R [1 ]
机构
[1] Childrens Hosp, Div Cardiol, Cardiovasc Device Dev Lab, Denver, CO 80218 USA
[2] Univ Colorado, MEDICA Res Ctr, Dept Mech Engn, Boulder, CO 80309 USA
关键词
heart valve prostheses; haemodynamics; spectral analysis; implantable sensors;
D O I
10.1007/BF02348084
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The development of a 'smart' heart valve prosthesis, with the intrinsic ability to monitor thrombus formation, mechanical failure and local haemodynamics and to relay this information externally, would be of significant help to clinicians. The first step towards such a valve is development of the sensors and examination of whether sensor output provides predictive information on function. Custom-made piezo-electric sensors were mounted onto the housing of mechanical valves with various layers of simulated thrombus. and bioprosthetic valves with normal and stiffened leaflets. Sensor output was examined using joint time-frequency analysis. Sensors were able to detect leaflet opening and closing with high fidelity for all types of valve. The frequency content of the closing sounds for the mechanical valves contained several peaks between 100 Hz and 10 kHz, whereas closing sounds for the bioprosthetic valve contained energy in a lower frequency range (< 1 kHz). A frequency peak of 47 +/- 15 Hz was seen for the normal bioprosthetic valve, this peak increased to 115 +/- 12 Hz for the valve with visibly stiffened leaflets. Total low-frequency (80-3500 Hz) energy content diminished predictably with increasing levels of thrombus for the mechanical valves. Lastly, closing sound intensity correlated well with closing pressure dynamics (dp/dt) (y = 190x-443, r = 0.90), indicating that the sensors also provide information on haemodynamics. These studies provide initial evidence regarding the use of embedded sensors to detect prosthetic valve function. Efforts to encapsulate these sensors with telemetry into a custom valve are currently underway.
引用
收藏
页码:416 / 424
页数:9
相关论文
共 13 条
[1]   TRANSIENT-BEHAVIOR ANALYSIS OF A MECHANICAL MONOLEAFLET HEART-VALVE PROSTHESIS IN THE CLOSING PHASE [J].
CHEON, GJ ;
CHANDRAN, KB .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1994, 116 (04) :452-459
[2]  
de Mol BA, 2001, ARTIF ORGANS, V25, P63
[3]   COMPARISON OF SPECTRAL TECHNIQUES FOR COMPUTER-ASSISTED CLASSIFICATION OF SPECTRA OF HEART SOUNDS IN PATIENTS WITH PORCINE BIOPROSTHETIC VALVES [J].
DURAND, LG ;
GUO, Z ;
SABBAH, HN ;
STEIN, PD .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1993, 31 (03) :229-236
[4]  
Eberhardt A C, 1995, J Heart Valve Dis, V4, P649
[5]  
Erickson R L, 1994, J Heart Valve Dis, V3, P330
[6]   VIBRATIONAL ANALYSIS OF BIOPROSTHETIC HEART-VALVE LEAFLETS USING NUMERICAL-MODELS - EFFECTS OF LEAFLET STIFFENING, CALCIFICATION, AND PERFORATION [J].
HAMID, MS ;
SABBAH, HN ;
STEIN, PD .
CIRCULATION RESEARCH, 1987, 61 (05) :687-694
[7]  
LEVY RJ, 1986, CRIT REV BIOCOMPAT, V2, P147
[8]   ACOUSTIC CLASSIFICATION OF THE STATE OF ARTIFICIAL-HEART VALVES [J].
PLEMONS, TD ;
HOVENGA, M .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 97 (04) :2326-2333
[9]   DYNAMIC ANALYSIS OF FLUTTER IN DISK TYPE MECHANICAL HEART-VALVE PROSTHESES [J].
PRABHU, AA ;
HWANG, NHC .
JOURNAL OF BIOMECHANICS, 1988, 21 (07) :585-590
[10]   DETECTION OF MECHANICAL CHANGES TO PROSTHETIC HEART-VALVES BY SPECTRAL-ANALYSIS OF VALVE CLOSING SOUNDS [J].
REYNOLDS, KJ ;
STEPHEN, RO .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 98 (01) :60-68