The development of an on-chip micro-flow injection analysis of nitrate with a cadmium reductor

被引:50
作者
Petsul, PH [1 ]
Greenway, GM [1 ]
Haswell, SJ [1 ]
机构
[1] Univ Hull, Dept Chem, Kingston Upon Hull HU6 7RX, N Humberside, England
关键词
mu FIA; nitrate; electro-osmotic flow; copperised cadmium;
D O I
10.1016/S0003-2670(00)01244-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of a method to determine nitrate by micro-flow injection analysis (mu FIA) is described, using electro-osmotic flow (EOF) control in a micro-chip with 300 mum wide and 115 mum deep channel wet etched in borosilicate glass. An in situ miniature copperised cadmium reductor column was produced by immobilisation within a micro-porous silica frit made from 10% formamide and 21% silicon dioxide. The reductor frit was located in the sample introduction reservoir of the device. Nitrate standards were placed in the sampling reservoir and an applied voltage of 50V relative to a waste reservoir was used to move the nitrate ions through the reductor frit for mu FIA. Silica frits without copperised cadmium were also introduced into the interconnecting channels of the mu FIA device to promote EOF and minimise hydrodynamic movement of solutions. The nitrite produced on leaving the reductor frit was detected using the Griess diazo-coupling reaction. The reaction was monitored in situ using a micro-spectrometer. A calibration was obtained between 0.5 and 20 muM with a correlation coefficient of 0.985. The R.S.D. at 5 muM NO(3)(-) was 8.3% (n = 6) and the limit of detection (3 sigma) was 0.51 muM (0.026 mug ml(-1)) NO(3)(-). (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:155 / 161
页数:7
相关论文
共 13 条
[1]   SIMULTANEOUS FLOW-INJECTION DETERMINATION OF NITRATE AND NITRITE IN WATER BY GAS-PHASE CHEMILUMINESCENCE [J].
AOKI, T ;
WAKABAYASHI, M .
ANALYTICA CHIMICA ACTA, 1995, 308 (1-3) :308-312
[2]   Microstructure for efficient continuous flow mixing [J].
Bessoth, FG ;
deMello, AJ ;
Manz, A .
ANALYTICAL COMMUNICATIONS, 1999, 36 (06) :213-215
[3]   Sequential injection sandwich technique for the simultaneous determination of nitrate and nitrite [J].
Cerdà, A ;
Oms, MT ;
Forteza, R ;
Cerdà, V .
ANALYTICA CHIMICA ACTA, 1998, 371 (01) :63-71
[4]   The fabrication of micro-porous silica structures for micro-reactor technology [J].
Christensen, PD ;
Johnson, SWP ;
McCreedy, T ;
Skelton, V ;
Wilson, NG .
ANALYTICAL COMMUNICATIONS, 1998, 35 (10) :341-343
[5]   Dissolved iron at subnanomolar levels in the Southern Ocean as determined by ship-board analysis [J].
de Jong, JTM ;
den Das, J ;
Bathmann, U ;
Stoll, MHC ;
Kattner, G ;
Nolting, RF ;
de Baar, HJW .
ANALYTICA CHIMICA ACTA, 1998, 377 (2-3) :113-124
[6]   Simultaneous determination of nitrite and nitrate in various samples using flow injection with spectrophotometric detection [J].
Ensafi, AA ;
Kazemzadeh, A .
ANALYTICA CHIMICA ACTA, 1999, 382 (1-2) :15-21
[7]  
FIEHN H, 1995, MESA MG, P289
[8]   Theoretical considerations of chemical reactions in micro-reactors operating under electroosmotic and electrophoretic control [J].
Fletcher, PDI ;
Haswell, SJ ;
Paunov, VN .
ANALYST, 1999, 124 (09) :1273-1282
[9]   Characterisation of a micro-total analytical system for the determination of nitrite with spectrophotometric detection [J].
Greenway, GM ;
Haswell, SJ ;
Petsul, PH .
ANALYTICA CHIMICA ACTA, 1999, 387 (01) :1-10
[10]   Column preconcentration analysis-spectrophotometric determination of nitrate and nitrite by a diazotization-coupling reaction [J].
Horita, K ;
Wang, GF ;
Satake, M .
ANALYST, 1997, 122 (12) :1569-1574