Fiber-optic probe for noninvasive real-time determination of tissue optical properties at multiple wavelengths

被引:106
作者
Dam, JS [1 ]
Pedersen, CB
Dalgaard, T
Fabricius, PE
Aruna, P
Andersson-Engels, S
机构
[1] Bang & Olufsen Medicom AS, DK-7600 Struer, Denmark
[2] Lund Inst Technol, Dept Phys, SE-22100 Lund, Sweden
[3] Anna Univ, Madras 600025, Tamil Nadu, India
关键词
D O I
10.1364/AO.40.001155
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We present a compact, fast, and versatile fiber-optic probe system for real-time determination of tissue optical properties from spatially resolved continuous-wave diffuse reflectance measurements. The system collects one set of reflectance data from six source-detector distances at four arbitrary wavelengths with a maximum overall sampling rate of 100 Hz. Multivariate calibration techniques based on two-dimensional polynomial fitting are employed to extract and display the absorption and reduced scattering coefficients in real-time mode. The four wavelengths of the current configuration are 660, 785, 805, and 974 nm, respectively. Cross-validation tests on a 6 x 7 calibration matrix of Intralipid-dye phantoms showed that the mean prediction error at, e.g., 785 nm was 2.8% for the absorption coefficient and 1.3% for the reduced scattering coefficient. The errors are relative to the range of the optical properties of the phantoms at 785 nm, which were 0-0.3/cm for the absorption coefficient and 6-16/cm for the reduced scattering coefficient. Finally, we also present and discuss results from preliminary skin tissue measurements. (C) 2001 Optical Society of America