The creep and wear of highly cross-linked polyethylene - A three-year randomised, controlled trial using radiostereometric analysis

被引:121
作者
Glyn-Jones, S. [1 ]
McLardy-Smith, P. [1 ]
Gill, H. S. [1 ]
Murray, D. W. [1 ]
机构
[1] Nuffield Orthopaed Ctr, Oxford OX3 7LD, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2008年 / 90B卷 / 05期
关键词
D O I
10.1302/0301-620X.90B5.20545
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis. We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years. After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly crosslinked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/ yr, SD 0.06; standard 0.07 mm/ yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial. We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.
引用
收藏
页码:556 / 561
页数:6
相关论文
共 42 条
[1]
Multidirectional deformation in fully congruent acetabular components [J].
Akisue, T ;
Bauer, TW ;
Yamaguchi, M ;
Matejczyk, MB ;
Stulberg, BN ;
Wilde, AH .
JOURNAL OF ARTHROPLASTY, 1999, 14 (08) :1011-1018
[2]
Should total hip arthroplasty femoral components be designed to subside?: A radiostereometric analysis study of the Charnley Elite and Exeter stems [J].
Alfaro-Adrián, J ;
Gill, HS ;
Murray, DW .
JOURNAL OF ARTHROPLASTY, 2001, 16 (05) :598-606
[3]
Altman DG., 1990, PRACTICAL STAT MED R
[4]
Archibeck M J, 2001, Instr Course Lect, V50, P185
[5]
Baker DA, 1999, J BIOMED MATER RES, V46, P573, DOI 10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO
[6]
2-A
[7]
The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene [J].
Baker, DA ;
Bellare, A ;
Pruitt, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2003, 66A (01) :146-154
[8]
Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners [J].
Bradford, L ;
Baker, DA ;
Graham, J ;
Chawan, A ;
Ries, MD ;
Pruitt, LA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) :1271-1282
[9]
Third-body wear of highly cross-linked polyethylene in a hip simulator [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
O'Connor, DO ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :553-561
[10]
Highly cross-linked polyethylene in cemented THA -: Randomized study of 61 hips [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :126-138