Measurement of small inter-scan fluctuations in voxel dimensions in magnetic resonance images using registration

被引:29
作者
Lemieux, L [1 ]
Barker, GJ [1 ]
机构
[1] UCL, Inst Neurol, Dept Clin Neurol, Epilepsy Res Grp, London WC1N 3BG, England
关键词
MRI; geometric distortion; QA; image registration;
D O I
10.1118/1.598288
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present a method to estimate fluctuations in voxel dimensions in MR images. In the proposed method images of a standard test object are acquired serially and registered to an initial dataset, We present the results of experiments where changes in voxel dimensions between 0.1% and 2% were prescribed and measured with an accuracy of 0.02%, The method can be integrated as part of a standard MRI Quality Assurance program and may also be useful to correct for inter-scan variability introduced by fluctuations in gradient performance. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02306-2].
引用
收藏
页码:1049 / 1054
页数:6
相关论文
共 15 条
[1]  
[Anonymous], 1988, MAGN RESON IMAGING, V6, P195
[2]   SEMIAUTOMATED QUALITY ASSURANCE FOR QUANTITATIVE MAGNETIC-RESONANCE-IMAGING [J].
BARKER, GJ ;
TOFTS, PS .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :585-595
[3]   Accurate registration of serial 3D MR brain images and its application to visualizing change in neurodegenerative disorders [J].
Freeborough, PA ;
Woods, RP ;
Fox, NC .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (06) :1012-1022
[4]   Spatial registration and normalization of images [J].
Friston, KJ ;
Ashburner, J ;
Frith, CD ;
Poline, JB ;
Heather, JD ;
Frackowiak, RSJ .
HUMAN BRAIN MAPPING, 1995, 3 (03) :165-189
[5]   A REGISTRATION AND INTERPOLATION PROCEDURE FOR SUBVOXEL MATCHING OF SERIALLY ACQUIRED MR-IMAGES [J].
HAJNAL, JV ;
SAEED, N ;
SOAR, EJ ;
OATRIDGE, A ;
YOUNG, IR ;
BYDDER, GM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1995, 19 (02) :289-296
[6]   MRI SCANNER VARIABILITY STUDIES USING A SEMIAUTOMATED ANALYSIS SYSTEM [J].
HYDE, RJ ;
ELLIS, JH ;
GARDNER, EA ;
ZHANG, YT ;
CARSON, PL .
MAGNETIC RESONANCE IMAGING, 1994, 12 (07) :1089-1097
[7]   A PATIENT-TO-COMPUTED-TOMOGRAPHY IMAGE REGISTRATION METHOD BASED ON DIGITALLY RECONSTRUCTED RADIOGRAPHS [J].
LEMIEUX, L ;
JAGOE, R ;
FISH, DR ;
KITCHEN, ND ;
THOMAS, DGT .
MEDICAL PHYSICS, 1994, 21 (11) :1749-1760
[8]  
LEMIEUX L, 1996, P ISMRM 4 SCI M NEW, P600
[9]  
LEMIEUX L, 1997, RADIOLOGY P, V205, P534
[10]   QUALITY ASSURANCE METHODS AND PHANTOMS FOR MAGNETIC-RESONANCE-IMAGING - REPORT OF AAPM NUCLEAR-MAGNETIC-RESONANCE TASK GROUP NO-1 [J].
PRICE, RR ;
AXEL, L ;
MORGAN, T ;
NEWMAN, R ;
PERMAN, W ;
SCHNEIDERS, N ;
SELIKSON, M ;
WOOD, M ;
THOMAS, SR .
MEDICAL PHYSICS, 1990, 17 (02) :287-295