A PATIENT-TO-COMPUTED-TOMOGRAPHY IMAGE REGISTRATION METHOD BASED ON DIGITALLY RECONSTRUCTED RADIOGRAPHS

被引:169
作者
LEMIEUX, L [1 ]
JAGOE, R [1 ]
FISH, DR [1 ]
KITCHEN, ND [1 ]
THOMAS, DGT [1 ]
机构
[1] ROYAL POSTGRAD MED SCH,DEPT MED PHYS,LONDON W12 0NN,ENGLAND
关键词
D O I
10.1118/1.597276
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An automatic method for the accurate registration of computed tomography (CT) data with two camera-calibrated radiographs is presented. The registration is based on the skull as visualized both in the plain radiographs and in radiographs digitally reconstructed from CT. A reference coordinate system is established based on the radiographic projection parameters obtained using an angiographic stereotactic localizer. The CT-derived reconstructed radiographs are aligned iteratively at multiple resolutions until a best match is found by adjusting the position and orientation of the CT data set relative to the reference coordinate system. The results of experiments with a skull phantom performed under stereotactic control which show that reliable registration is possible with an accuracy better than 1 mm are presented. Possible applications include intraoperative patient-to-CT frameless registration and registration of radiographic data with frameless CT for depth electroencephalogram electrode position confirmation. © 1994, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:1749 / 1760
页数:12
相关论文
共 18 条
[1]   USE OF A FRAMELESS, ARMLESS STEREOTAXIC WAND FOR BRAIN-TUMOR LOCALIZATION WITH 2-DIMENSIONAL AND 3-DIMENSIONAL NEUROIMAGING [J].
BARNETT, GH ;
KORMOS, DW ;
STEINER, CP ;
WEISENBERGER, J .
NEUROSURGERY, 1993, 33 (04) :674-678
[2]   3-DIMENSIONAL VERIFICATION OF PATIENT PLACEMENT DURING RADIOTHERAPY USING PORTAL IMAGES [J].
BIJHOLD, J .
MEDICAL PHYSICS, 1993, 20 (02) :347-356
[3]  
GALLOWAY RL, 1990, CRIT REV BIOMED ENG, V18, P181
[4]   INTERACTIVE IMAGE-GUIDED NEUROSURGERY [J].
GALLOWAY, RL ;
MACIUNAS, RJ ;
EDWARDS, CA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (12) :1226-1231
[5]   AUTOMATIC ONLINE INSPECTION OF PATIENT SETUP IN RADIATION-THERAPY USING DIGITAL PORTAL IMAGES [J].
GILHUIJS, KGA ;
VANHERK, M .
MEDICAL PHYSICS, 1993, 20 (03) :667-677
[6]   MULTIMODALITY IMAGE INTEGRATION FOR STEREOTAXIC SURGICAL PLANNING [J].
HENRI, CJ ;
COLLINS, DL ;
PETERS, TM .
MEDICAL PHYSICS, 1991, 18 (02) :167-177
[7]   MEDICAL IMAGE REGISTRATION USING KNOWLEDGE OF ADJACENCY OF ANATOMICAL STRUCTURES [J].
HILL, DLG ;
HAWKES, DJ .
IMAGE AND VISION COMPUTING, 1994, 12 (03) :173-178
[8]   A FRAMELESS, ARMLESS NAVIGATIONAL SYSTEM FOR COMPUTER-ASSISTED NEUROSURGERY [J].
KATO, A ;
YOSHIMINE, T ;
HAYAKAWA, T ;
TOMITA, Y ;
IKEDA, T ;
MITOMO, M ;
HARADA, K ;
MOGAMI, H .
JOURNAL OF NEUROSURGERY, 1991, 74 (05) :845-849
[9]  
KITCHEN ND, 1993, IN PRESS STEREOTACTI
[10]   VOXEL-BASED LOCALIZATION IN FRAME-BASED AND FRAMELESS STEREOTAXY AND ITS ACCURACY [J].
LEMIEUX, L ;
KITCHEN, ND ;
HUGHES, SW ;
THOMAS, DGT .
MEDICAL PHYSICS, 1994, 21 (08) :1301-1310