Performance and safety of an integrated bihormonal artificial pancreas for fully automated glucose control at home

被引:81
作者
Blauw, H. [1 ,2 ]
van Bon, A. C. [3 ]
Koops, R. [2 ]
DeVries, J. H. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Endocrinol, NL-1105 AZ Amsterdam, Netherlands
[2] Inreda Diabet BV, Goor, Netherlands
[3] Rijnstate Hosp, Arnhem, Netherlands
关键词
clinical trial; glucagon; glycaemic control; insulin therapy; type; 1; diabetes; RANDOMIZED CROSSOVER TRIAL; FREE-LIVING CONDITIONS; INSULIN PUMP THERAPY; CLOSED-LOOP SYSTEM; GLYCEMIC CONTROL; BIONIC PANCREAS; TYPE-1; HYPOGLYCEMIA; FREQUENCY; AWARENESS;
D O I
10.1111/dom.12663
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Aims: To assess the performance and safety of an integrated bihormonal artificial pancreas system consisting of one wearable device and two wireless glucose sensor transmitters during short-term daily use at home. Methods: Adult patients with type 1 diabetes using an insulin pump were invited to enrol in this randomized crossover study. Treatment with the artificial pancreas started with a day and night in the clinical research centre, followed by 3 days at home. The control period consisted of 4 days of insulin pump therapy at home with blinded continuous glucose monitoring for data collection. Days 2-4 were predefined as the analysis period, with median glucose as the primary outcome. Results: A total of 10 patients completed the study. The median [interquartile range (IQR)] glucose level was similar for the two treatments [7.3 (7.0-7.6) mmol/l for the artificial pancreas vs. 7.7 (7.0-9.0) mmol/l for the control; p= 0.123]. The median (IQR) percentage of time spent in euglycaemia (3.9-10 mmol/l) was longer during use of the artificial pancreas [84.7 (82.2-87.8)% for the artificial pancreas vs. 68.5 (57.9-83.6)% for the control; p= 0.007]. Time in hypoglycaemia was 1.3 (0.2-3.2)% for the artificial pancreas and 2.4 (0.4-10.3)% for the control treatment (p= 0.139). Separate analysis of daytime and night-time showed that the improvements were mainly achieved during the night. Conclusions: The results of this pilot study suggest that our integrated artificial pancreas provides better glucose control than insulin pump therapy in patients with type 1 diabetes at home and that the treatment is safe.
引用
收藏
页码:671 / 677
页数:7
相关论文
共 18 条
[1]
Pharmacokinetics and pharmacodynamics of various glucagon dosages at different blood glucose levels [J].
Blauw, H. ;
Wendl, I. ;
DeVries, J. H. ;
Heise, T. ;
Jax, T. .
DIABETES OBESITY & METABOLISM, 2016, 18 (01) :34-39
[2]
Carroll Mary F, 2005, Endocr Pract, V11, P55
[3]
Statistical Tools to Analyze Continuous Glucose Monitor Data [J].
Clarke, William ;
Kovatchev, Boris .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 :S45-S54
[4]
REDUCED AWARENESS OF HYPOGLYCEMIA IN ADULTS WITH IDDM - A PROSPECTIVE-STUDY OF HYPOGLYCEMIC FREQUENCY AND ASSOCIATED SYMPTOMS [J].
CLARKE, WL ;
COX, DJ ;
GONDERFREDERICK, LA ;
JULIAN, D ;
SCHLUNDT, D ;
POLONSKY, W .
DIABETES CARE, 1995, 18 (04) :517-522
[5]
Efficacy of dual-hormone artificial pancreas to alleviate the carbohydrate-counting burden of type 1 diabetes: A randomized crossover trial [J].
Gingras, V. ;
Rabasa-Lhoret, R. ;
Messier, V. ;
Ladouceur, M. ;
Legault, L. ;
Haidar, A. .
DIABETES & METABOLISM, 2016, 42 (01) :47-54
[6]
FREQUENCY OF SEVERE HYPOGLYCEMIA IN PATIENTS WITH TYPE-1 DIABETES WITH IMPAIRED AWARENESS OF HYPOGLYCEMIA [J].
GOLD, AE ;
MACLEOD, KM ;
FRIER, BM .
DIABETES CARE, 1994, 17 (07) :697-703
[7]
Comparison of dual-hormone artificial pancreas, single-hormone artificial pancreas, and conventional insulin pump therapy for glycaemic control in patients with type 1 diabetes: an open-label randomised controlled crossover trial [J].
Haidar, Ahmad ;
Legault, Laurent ;
Messier, Virginie ;
Mitre, Tina Maria ;
Leroux, Catherine ;
Rabasa-Lhoret, Remi .
LANCET DIABETES & ENDOCRINOLOGY, 2015, 3 (01) :17-26
[8]
Reimbursement for Continuous Glucose Monitoring [J].
Heinemann, Lutz ;
DeVries, J. Hans .
DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 :48-52
[9]
Stable Liquid Glucagon Formulations for Rescue Treatment and Bi-Hormonal Closed-Loop Pancreas [J].
Jackson, Melanie A. ;
Caputo, Nicholas ;
Castle, Jessica R. ;
David, Larry L. ;
Roberts, Charles T., Jr. ;
Ward, W. Kenneth .
CURRENT DIABETES REPORTS, 2012, 12 (06) :705-710
[10]
Automated Control of an Adaptive Bihormonal, Dual-Sensor Artificial Pancreas and Evaluation During Inpatient Studies [J].
Jacobs, Peter G. ;
El Youssef, Joseph ;
Castle, Jessica ;
Bakhtiani, Parkash ;
Branigan, Deborah ;
Breen, Matthew ;
Bauer, David ;
Preiser, Nicholas ;
Leonard, Gerald ;
Stonex, Tara ;
Ward, W. Kenneth .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (10) :2569-2581