Evaluation of a direct injection high-efficiency nebulizer (DIHEN) by comparison with a high-efficiency nebulizer (HEN) coupled to a cyclonic spray chamber as a liquid sample introduction system for ICP-AES

被引:60
作者
Todolí, JL [1 ]
Mermet, JM
机构
[1] Univ Alicante, Dept Quim Analit, E-03080 Alicante, Spain
[2] Univ Lyon 1, Sci Analyt Lab, UMR 5619, F-69622 Villeurbanne, France
关键词
D O I
10.1039/b009430g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the present work the advantages and limitations of introducing the sample directly into the plasma through the use of a direct injection high-efficiency nebulizer (DIHEN) have been evaluated in ICP-AES. The reference system was a high-efficiency nebulizer (HEN), precursor of the DIHEN, coupled to a cyclonic spray chamber. It was found that the plasma can only be operated under non-robust conditions because of the high solvent loading. The issues considered in the present work were: (i) the aerosol characteristics; (ii) the ICP-AES analytical figures of merit; and (iii) the interferences induced by the presence of inorganic as well as organic matrices. The results indicated that the aerosols obtained by the DIHEN are coarser than those leaving the cyclonic spray chamber. Nonetheless, and due to the higher analyte transport efficiency, the DIHEN affords ICP-AES sensitivities and LODs up to 10-18 times better than the cyclonic spray chamber. Also, it was found that, under the plasma conditions tested, the optimum nebulizer gas and liquid flow rates in terms of sensitivity for the DIHEN were very different from those for the HEN with the cyclonic spray chamber. In the first case, the signal peaked at rather low liquid (Q(l)) and gas (Q(g)) flow rates, whereas for the cyclonic spray chamber moderate and/or high liquid and gas flow rates were advisable. Under a given set of plasma conditions, the DIHEN has been proven to produce problems due to deterioration in the plasma, derived from the higher solvent load with respect to the HEN coupled to the cyclonic spray chamber. Despite these findings and the fact that a non-robust plasma was used, the matrix effects for inorganic and organic species were significantly minimized, for a low Q(l) and for some lines, in the case of the DIHEN.
引用
收藏
页码:514 / 520
页数:7
相关论文
共 44 条
[1]   A large bore-direct injection high efficiency nebulizer for inductively coupled plasma spectrometry [J].
Acon, BW ;
McLean, JA ;
Montaser, A .
ANALYTICAL CHEMISTRY, 2000, 72 (08) :1885-1893
[2]   Ultratrace and isotope analysis of long-lived radionuclides by inductively coupled plasma quadrupole mass spectrometry using a direct injection high efficiency nebulizer [J].
Becker, JS ;
Dietze, HJ ;
McLean, JA ;
Montaser, A .
ANALYTICAL CHEMISTRY, 1999, 71 (15) :3077-3084
[3]   Application of double-focusing sector field ICP mass spectrometry with shielded torch using different nebulizers for ultratrace and precise isotope analysis of long-lived radionuclides - Invited lecture [J].
Becker, JS ;
Dietze, HJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (09) :1493-1500
[4]   Non-spectral interference effects in inductively coupled plasma mass spectrometry using direct injection high efficiency and microconcentric nebulisation [J].
Björn, E ;
Frech, W .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (01) :4-11
[5]   EXPERIMENTAL EVALUATION OF THE NUKIYAMA-TANASAWA EQUATION FOR PNEUMATIC NEBULIZERS USED IN PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
CANALS, A ;
HERNANDIS, V ;
BROWNER, RF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1990, 5 (01) :61-66
[6]   FUNDAMENTAL-STUDIES ON PNEUMATIC GENERATION AND AEROSOL TRANSPORT IN ATOMIC SPECTROMETRY - EFFECT OF MINERAL ACIDS ON EMISSION INTENSITY IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
CANALS, A ;
HERNANDIS, V ;
TODOLI, JL ;
BROWNER, RF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :305-321
[7]   Selection of the operating conditions for overcoming acids and sodium chloride non-spectroscopic interferences in inductively coupled plasma atomic emission spectrometry:: effect of the liquid-to-vapor mass ratios [J].
Catasús, M ;
Todolí, JL ;
Gras, L ;
Hernandis, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (09) :1203-1206
[8]   DIAGNOSTIC STUDIES ON DESOLVATED AEROSOLS FROM ULTRASONIC NEBULIZERS [J].
CLIFFORD, RH ;
SOHAL, P ;
LIU, HY ;
MONTASER, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (09) :1107-1122
[9]   Comparison of the effect of acetic acid with axially and radially viewed inductively coupled plasma atomic emission spectrometry: influence of the operating conditions [J].
Dubuisson, C ;
Poussel, E ;
Mermet, JM ;
Todoli, JL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (01) :63-67
[10]   Effect of sodium during the aerosol transport and filtering in inductively coupled plasma atomic emission spectrometry [J].
Dubuisson, C ;
Poussel, E ;
Todoli, JL ;
Mermet, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (04) :593-600