Severe intraprosthetic regurgitation by immobile leaflet after trans-catheter aortic valve implantation

被引:30
作者
Al-Attar, Nawwar [1 ]
Himbert, Dominique [2 ]
Vahanian, Alec [2 ]
Nataf, Patrick
机构
[1] Univ Paris 07, Hop Bichat, Serv Chirurg Cardiaque, Dept Cardiac Surg,AP HP, F-75877 Paris 18, France
[2] Univ Paris 07, Hop Bichat, Dept Cardiol, AP HP, F-75877 Paris 18, France
关键词
Trans-catheter valves; Complications; Aortic valve;
D O I
10.1016/j.ejcts.2010.08.007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aortic regurgitation mainly in the form of paravalvular leaks is a frequent complication of trans-catheter aortic valve implantation (TAVI). We describe a case of an 86-year-old woman with severe aortic stenosis, who underwent trans-apical TAVI with a 23-mm Edwards-SAPIEN valve. Immediately post-implantation, severe intravalvular leak was observed on trans-esophageal echocardiogram (TEE) due to an immobile cusp associated with left-ventricular distension and cardiovascular collapse. Despite successfully treating the leak with the implantation of a second valve of the same diameter within the first one, the hemodynamic status remained unstable. Peripheral extracorporeal membrane oxygenation (ECMO) was established but resulted in a fatal outcome due to intractable heart failure. (C) 2010 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:591 / 592
页数:2
相关论文
共 7 条
[1]   Unexpected Complications of Transapical Aortic Valve Implantation [J].
Al-Attar, Nawwar ;
Ghodbane, Walid ;
Himbert, Dominique ;
Rau, Cederic ;
Raffoul, Richard ;
Messika-Zeitoun, David ;
Brochet, Eric ;
Vahanian, Alec ;
Nataf, Patrick .
ANNALS OF THORACIC SURGERY, 2009, 88 (01) :90-94
[2]  
Al-Attar N, 2009, HERZ, V34, P367, DOI 10.1007/s00059-009-3259-9
[3]  
ELTCHANINOFF H, 2009, LATE BREAKING INTERV
[4]   Minimally Invasive Off-Pump Aortic Valve Implantation: The Surgical Safety Net [J].
Kempfert, Joerg ;
Walther, Thomas ;
Borger, Michael A. ;
Lehmann, Sven ;
Blumenstein, Johannes ;
Fassl, Jens ;
Schuler, Gerhard ;
Mohr, Friedrich-Wilhelm .
ANNALS OF THORACIC SURGERY, 2008, 86 (05) :1665-1668
[5]   Everted leaflet of a bovine pericardial aortic valve [J].
Orihashi, Kazumasa ;
Kurosaki, Tatsuya ;
Sueda, Taijiro .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2010, 10 (06) :1059-1060
[6]   Transcatheter Aortic Valve Implantation for the Treatment of Severe Symptomatic Aortic Stenosis in Patients at Very High or Prohibitive Surgical Risk Acute and Late Outcomes of the Multicenter Canadian Experience [J].
Rodes-Cabau, Josep ;
Webb, John G. ;
Cheung, Anson ;
Ye, Jian ;
Dumont, Eric ;
Feindel, Christopher M. ;
Osten, Mark ;
Natarajan, Madhu K. ;
Velianou, James L. ;
Martucci, Giuseppe ;
DeVarennes, Benoit ;
Chisholm, Robert ;
Peterson, Mark D. ;
Lichtenstein, Samuel V. ;
Nietlispach, Fabian ;
Doyle, Daniel ;
DeLarochelliere, Robert ;
Teoh, Kevin ;
Chu, Victor ;
Dancea, Adrian ;
Lachapelle, Kevin ;
Cheema, Asim ;
Latter, David ;
Horlick, Eric .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (11) :1080-1090
[7]   Trans-apical aortic valve implantation: univariate and multivariate analyses of the early results from the SOURCE registry [J].
Wendler, Olaf ;
Walther, Thomas ;
Nataf, Patrick ;
Rubino, Paolo ;
Schroefel, Holger ;
Thielmann, Matthias ;
Treede, Hendrik ;
Thomas, Martyn .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2010, 38 (02) :119-126