Measured lung dose correction factors for 50 MV photons

被引:9
作者
Blomquist, M [1 ]
Karlsson, M [1 ]
机构
[1] Umea Univ, Dept Radiat Phys, SE-90185 Umea, Sweden
关键词
D O I
10.1088/0031-9155/43/11/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Some clinically relevant measurements of lung tissue/water equivalent interfaces have been performed for a 50 MV therapeutic x-ray beam. The purpose was to investigate the severity of dose perturbation effects in lung tissue and adjacent tissues using an energy well above the common clinical practice in thoracic irradiations. The phantoms were constructed of solid water, PMMA and white polystyrene as soft tissue (water) equivalents, and cork was used as the lung tissue equivalent. Measurements were performed using radiographic film and a cylindrical ionization chamber. The results show that the degradation of the 20/80% beam penumbra in the lung region is severe, up to 2.5 times the penumbra in water for a 10 cm thick lung with a density of 0.30 x 10(3) kg m(-3). The lack of electronic equilibrium in the low-density region can cause underdosage at the lung/tumour interface of up to 30% of maximum target dose, and the build-up depth to 95% of target dose in unit density tissue behind the lung may be as large as 22 mm. It is also shown that these figures strongly depend on patient anatomy and beam size and why a careful calculation of the individual dose distribution is needed for optimal choice of photon beam energy in thoracic treatments.
引用
收藏
页码:3225 / 3234
页数:10
相关论文
共 18 条
[11]   THE INFLUENCE OF IONIZATION-CHAMBER AND PHANTOM DESIGN ON THE MEASUREMENT OF LUNG DOSE IN PHOTON BEAMS [J].
RICE, RK ;
HANSEN, JL ;
CHIN, LM ;
MIJNHEER, BJ ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1988, 15 (06) :884-890
[12]   DENSITY PATTERNS IN THE NORMAL LUNG AS DETERMINED BY COMPUTED-TOMOGRAPHY [J].
ROSENBLUM, LJ ;
MAUCERI, RA ;
WELLENSTEIN, DE ;
THOMAS, FD ;
BASSANO, DA ;
RAASCH, BN ;
CHAMBERLAIN, CC ;
HEITZMAN, ER .
RADIOLOGY, 1980, 137 (02) :409-416
[13]  
Satherberg A, 1996, MED PHYS, V23, P495, DOI 10.1118/1.597681
[14]   FILM DOSIMETRY IN WATER IN A 23-MV THERAPEUTIC PHOTON-BEAM [J].
VANBATTUM, LJ ;
HEIJMEN, BJM .
RADIOTHERAPY AND ONCOLOGY, 1995, 34 (02) :152-159
[15]   LUNG DENSITY AS MEASURED BY COMPUTERIZED-TOMOGRAPHY - IMPLICATIONS FOR RADIOTHERAPY [J].
VANDYK, J ;
KEANE, TJ ;
RIDER, WD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1982, 8 (08) :1363-1372
[16]   Comparison of dose homogeneity effects due to electron equilibrium loss in lung for 6 MV and 18 MV photons [J].
White, PJ ;
Zwicker, RD ;
Huang, DT .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (05) :1141-1146
[17]   Dosimetric considerations in radiation therapy of coin lesions of the lung [J].
Yorke, E ;
Harisladis, L ;
Wessels, B ;
Aghdam, H ;
Altemus, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (02) :481-487
[18]   DOSE CORRECTIONS FOR LOW-DENSITY TISSUE INHOMOGENEITIES AND AIR CHANNELS FOR 10-MV X-RAYS [J].
YOUNG, MEJ ;
KORNELSEN, RO .
MEDICAL PHYSICS, 1983, 10 (04) :450-455