Joule heating and heat transfer in poly(dimethylsiloxane) microfluidic systems

被引:255
作者
Erickson, D [1 ]
Sinton, D [1 ]
Li, DQ [1 ]
机构
[1] Univ Toronto, Dept Mech & Ind Engn, Toronto, ON M5S 3G8, Canada
关键词
D O I
10.1039/b306158b
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Joule heating is a significant problem in electrokinetically driven microfluidic chips, particularly polymeric systems where low thermal conductivities amplify the difficulty in rejecting this internally generated heat. In this work, a combined experimental (using a microscale thermometry technique) and numerical (using a 3D "whole- chip" finite element model) approach is used to examine Joule heating and heat transfer at a microchannel intersection in poly(dimethylsiloxane) (PDMS), and hybrid PDMS/Glass microfluidic systems. In general the numerical predictions and the experimental results agree quite well (typically within +/-3degreesC), both showing dramatic temperature gradients at the intersection. At high potential field strengths a nearly five fold increase in the maximum buffer temperature was observed in the PDMS/PDMS chips over the PDMS/Glass systems. The detailed numerical analysis revealed that the vast majority of steady state heat rejection is through lower substrate of the chip, which was significantly impeded in the former case by the lower thermal conductivity PDMS substrate. The observed higher buffer temperature also lead to a number of significant secondary effects including a near doubling of the volume flow rate. Simple guidelines are proposed for improving polymeric chip design and thereby extend the capabilities of these microfluidic systems.
引用
收藏
页码:141 / 149
页数:9
相关论文
共 47 条
[1]  
[Anonymous], 2001, BIOCHIP TECHNOLOGY
[2]   Control of flow direction in microfluidic devices with polyelectrolyte multilayers [J].
Barker, SLR ;
Ross, D ;
Tarlov, MJ ;
Gaitan, M ;
Locascio, LE .
ANALYTICAL CHEMISTRY, 2000, 72 (24) :5925-5929
[3]   Polymer microfluidic devices [J].
Becker, H ;
Locascio, LE .
TALANTA, 2002, 56 (02) :267-287
[4]   Finite element simulation of an electroosmotic-driven flow division at a T-junction of microscale dimensions [J].
Bianchi, F ;
Ferrigno, A ;
Girault, HH .
ANALYTICAL CHEMISTRY, 2000, 72 (09) :1987-1993
[5]   An integrated fluorescence detection system in poly(dimethylsiloxane) for microfluidic applications [J].
Chabinyc, ML ;
Chiu, DT ;
McDonald, JC ;
Stroock, AD ;
Christian, JF ;
Karger, AM ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 2001, 73 (18) :4491-4498
[6]   Transient liquid crystal thermometry of microfabricated PCR vessel arrays [J].
Chaudhari, AM ;
Woudenberg, TM ;
Albin, M ;
Goodson, KE .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 1998, 7 (04) :345-355
[7]   Sample stacking revisited: A personal perspective [J].
Chien, RL .
ELECTROPHORESIS, 2003, 24 (03) :486-497
[8]   Microfabricated devices in biotechnology and biochemical processing [J].
Chován, T ;
Guttman, A .
TRENDS IN BIOTECHNOLOGY, 2002, 20 (03) :116-122
[9]   Patterned delivery of immunoglobulins to surfaces using microfluidic networks [J].
Delamarche, E ;
Bernard, A ;
Schmid, H ;
Michel, B ;
Biebuyck, H .
SCIENCE, 1997, 276 (5313) :779-781
[10]   Rapid prototyping of microfluidic systems in poly(dimethylsiloxane) [J].
Duffy, DC ;
McDonald, JC ;
Schueller, OJA ;
Whitesides, GM .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :4974-4984