Electrospray ionization mass spectrometry of tetracycline, oxytetracycline, chlorotetracycline, minocycline, and methacycline

被引:51
作者
Kamel, AM [1 ]
Brown, PR
Munson, B
机构
[1] Pfizer Inc, Div Cent Res, Dept Drug Metab, Groton, CT 06340 USA
[2] Univ Rhode Isl, Dept Chem, Kingston, RI 02881 USA
[3] Univ Delaware, Dept Chem & Biochem, Newark, DE 19716 USA
关键词
D O I
10.1021/ac9807114
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The effects of mobile-phase additives and analyte concentration on electrospray ionization mass spectra of a series of tetracyclines were investigated in both positive and negative ion modes. Only [M + H](+) and [M - H](-) ions were observed. The greatest sensitivity as [M + H](+) ions was obtained with 1% acetic acid and the greatest sensitivity as [M - H](-) ions was obtained using 50 mM. ammonium hydroxide. Sensitivities in the positive ion mode were greater than those in the negative ion mode. The sensitivity as [M + H](+) showed no systematic variation with pH; however, the sensitivity as [M - H](-) did increase with increasing pH. A larger linear range was observed for [M - H](-) than for [M + H](+) ions. Both [M + Na](+) and [IM + H](+) ions were observed with 0.5 mM sodium acetate and sodium iodide, but no adduct ions were observed with ammonium acetate. Some M2H+ ions were observed at higher concentrations. Cluster ions, Na(NaOAc)(n)(+) or Na(NaI)(n)(+), but no sample ions were observed using 5 mM salts. The data suggest that mechanisms in addition to solution ionization are involved in the formation of the ESI sample ions. The utility of mobile phases containing 1% HOAc or 50 mM NH4OH was demonstrated for chromatographic separations.
引用
收藏
页码:968 / 977
页数:10
相关论文
共 56 条