Long electrodes for radio frequency ablation: comparative study of surface versus intramural application

被引:5
作者
Berjano, EJ
Hornero, F
Atienza, F
Montero, A
机构
[1] Univ Politecn Valencia, Dept Ingn Elect, Valencia 46022, Spain
[2] Hosp Gen Univ Valencia, Dept Cardiol & Cirugia Cardiac, Valencia, Spain
关键词
ablation; atrial fibrillation; cardiac ablation; edge effect; electrode design; long electrode; radio frequency ablation;
D O I
10.1016/S1350-4533(03)00125-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is increasing use of radio frequency (R-F) ablation with long electrodes in the intraoperative treatment of atrial fibrillation. Nevertheless, the disparity in the lesion geometry in both depth and width is the major pitfall in the use of RF currents. The objective of this study was to differentiate the shape and size of long lesions created by three surface application electrodes (SAE) and two intramural electrodes (IE). The SAE included a standard multi-polar catheter, and two standard electrosurgical pencils. The IE consisted of a needle and a wire both intramurally buried. The lesions were created on fresh fragments of porcine ventricular tissue. The IE created lesions with a curved prism-like shape around the electrode body, with homogeneous characteristics along the lesion trajectory. On the contrary, the lesions created with the SAE were in the shape of an hourglass. They showed a different geometry between the central zone and the edge zone (p < 0.001 for depth and surface width). Electrical impedance evolution was recorded during the RF heating. We observed a slow decrease of the impedance in all the electrodes, except in the wire electrode. In conclusion, the results suggest that the IE might be a more suitable option than SAE when it is necessary to create long and homogeneous thermal lesions. (C) 2003 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:869 / 877
页数:9
相关论文
共 38 条
[21]   Endocardial and epicardial radiofrequency ablation in the treatment of atrial fibrillation with a new intra-operative device [J].
Melo, J ;
Adragao, P ;
Neves, J ;
Ferreira, M ;
Timóteo, A ;
Santiago, T ;
Ribeiras, R ;
Canada, M .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2000, 18 (02) :182-186
[22]   Morphological and physiological characteristics of discontinuous linear atrial ablations during atrial pacing and atrial fibrillation [J].
Mitchell, MA ;
McRury, ID ;
Everett, TH ;
Li, H ;
Mangrum, JM ;
Haines, DE .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (03) :378-386
[23]   Curative treatment of atrial fibrillation with intraoperative radiofrequency ablation: Short-term and midterm results [J].
Mohr, FW ;
Fabricius, AM ;
Falk, V ;
Autschbach, R ;
Doll, N ;
von Oppell, U ;
Diegeler, A ;
Kottkamp, H ;
Hindricks, G .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 123 (05) :919-927
[24]   COMPARISON OF IN-VIVO TISSUE TEMPERATURE PROFILE AND LESION GEOMETRY FOR RADIOFREQUENCY ABLATION WITH A SALINE-IRRIGATED ELECTRODE VERSUS TEMPERATURE CONTROL IN A CANINE THIGH MUSCLE PREPARATION [J].
NAKAGAWA, H ;
YAMANASHI, WS ;
PITHA, JV ;
ARRUDA, M ;
WANG, XZ ;
OHTOMO, K ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
LAZZARA, R ;
JACKMAN, WM .
CIRCULATION, 1995, 91 (08) :2264-2273
[25]   DEVELOPMENT OF A NEW INTRAOPERATIVE RADIOFREQUENCY ABLATION TECHNIQUE USING A NEEDLE ELECTRODE [J].
OHTAKE, H ;
MISAKI, T ;
MATSUNAGA, Y ;
WATANABE, G ;
TAKAHASHI, M ;
MATSUMOTO, I ;
KAWASUJI, M ;
WATANABE, Y .
ANNALS OF THORACIC SURGERY, 1994, 58 (03) :750-753
[26]   MEASUREMENTS OF TEMPERATURE DISTRIBUTIONS AT ELECTRO-SURGICAL DISPERSIVE ELECTRODE SITES [J].
OVERMYER, KM ;
PEARCE, JA ;
DEWITT, DP .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1979, 101 (01) :66-72
[27]   Radiofrequency multielectrode catheter ablation in the atrium [J].
Panescu, D ;
Fleischman, SD ;
Whayne, JG ;
Swanson, DK ;
Mirotznik, MS ;
McRury, I ;
Haines, DE .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (04) :899-915
[28]   3-DIMENSIONAL FINITE-ELEMENT ANALYSIS OF CURRENT-DENSITY AND TEMPERATURE DISTRIBUTIONS DURING RADIOFREQUENCY ABLATION [J].
PANESCU, D ;
WHAYNE, JG ;
FLEISCHMAN, SD ;
MIROTZNIK, MS ;
SWANSON, DK ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (09) :879-890
[29]   Intraoperative radiofrequency microbipolar coagulation to replace incisions of maze III procedure for correcting atrial fibrillation in patients with rheumatic valvular disease [J].
Patwardhan, AM ;
Dave, HH ;
Tamhane, AA ;
Pandit, SP ;
Dalvi, BV ;
Golam, K ;
Kaul, A ;
Chaukar, AP .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1997, 12 (04) :627-633
[30]  
PEARCE JA, 1986, ELECTROSURGERY, P17