Determination of pibutidine metabolites in human plasma by LC-MS/MS

被引:25
作者
Kato, K [1 ]
Jingu, S [1 ]
Ogawa, N [1 ]
Higuchi, S [1 ]
机构
[1] Taisho Pharmaceut Co Ltd, Pharmaceut Res Labs, Drug Metab Lab, Omiya, Saitama 3308530, Japan
关键词
pibutidine; metabolism; human plasma; LC-MS/MS; electrospray ionization; isotope effect;
D O I
10.1016/S0731-7085(00)00398-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel metabolite-screening procedure for pibutidine. an H-2-receptor antagonist, which uses high-performance liquid chromatography/tandem mass spectrometry (LC-MS/MS), demonstrated the presence of pibutidine and its four metabolites in plasma from volunteers who received a single dose of pibutidine hydrochloride. In order to quantitatively examine the metabolism of pibutidine, an assay based on LC-MS/MS was subsequently developed for the simultaneous determination of its metabolites in human plasma. Target analytes consisted of M-5, M-7 and M-8, which were prominently detected by the screening procedure, and M-9, which has pharmacological activity as an H-2-receptor antagonist. Metabolites and their deuterated internal standards were extracted from human plasma using an Oasis HLB extraction cartridge, and chromatographed on a Monitor C18M column. No isotope effects on chromatographic retention time were observed for any deuterated compounds, which were ionized using an electrospray ionization (ESI)- interface and detected by MS/MS in the selected reaction-monitoring (SRM) mode simultaneously with the corresponding metabolites. The assay was validated over the concentration range of 0.1 to 25.6 ng ml(-1) and used to determine the plasma levels of metabolites in volunteers following oral administration of a 20-mg dose of pibutidine hydrochloride. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:237 / 249
页数:13
相关论文
共 15 条
[1]  
ARAI M, 1999, JAP PHARM THER, V27, P619
[2]  
ARAI M, 1999, JAP PHARM THER, V27, P605
[3]  
AZUMA J, 1999, JPN PHARMACOL THER, V27, P973
[4]  
Bouvier ESP, 1997, LC GC-MAG SEP SCI, V15, P152
[5]  
Isobe Y, 1999, JPN PHARMACOL THER, V27, P543
[6]  
ISOBE Y, 1999, JPN PHARMACOL THER, V27, P261
[7]  
Kato K, 1999, RAPID COMMUN MASS SP, V13, P1626
[8]   Development and validation of a liquid chromatographic-tandem mass spectrometric method for the determination of pibutidine in human urine [J].
Kato, K ;
Jingu, S ;
Ogawa, N ;
Higuchi, S .
JOURNAL OF CHROMATOGRAPHY B, 2000, 740 (02) :187-193
[9]  
KIUCHI Y, 1999, JPN PHARMACOL THER, V27, P289
[10]   Development of high-performance liquid chromatography tandem mass spectrometric methods for the determination of a new oxytocin receptor antagonist (L-368,899) extracted from human plasma and urine: a case of lack of specificity due to the presence of metabolites [J].
Matuszewski, BK ;
Chavez-Eng, CM ;
Constanzer, ML .
JOURNAL OF CHROMATOGRAPHY B, 1998, 716 (1-2) :195-208