The measurement of creep in ultrahigh molecular weight polyethylene - A comparison of conventional versus highly cross-linked polyethylene

被引:37
作者
Estok, DM [1 ]
Bragdon, CR [1 ]
Plank, GR [1 ]
Huang, A [1 ]
Muratoglu, OK [1 ]
Harris, WH [1 ]
机构
[1] Massachusetts Gen Hosp, Orthopaed BIomech & Biomat Lab, Boston, MA 02114 USA
关键词
polyethylene; cross-linking; creep; hip; arthroplasty;
D O I
10.1016/j.arth.2004.08.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Quantification of creep of highly cross-linked polyethylene would enable separation of creep from wear when evaluating femoral head penetration into polyethylene. We compared creep magnitude of a highly cross-linked versus conventional polyethylene in the laboratory. Twelve acetabular liners of each material were tested, 6 of which had a 32-mm inner diameter (ID) and 6 had 28-mm 11). Creep was measured using coordinate measuring machines during loading at 2 Hz without motion to 4 million cycles. Penetration into 32-mm ID conventional liners reached 97 mum versus 107 pm for highly cross-linked material, not significant. Penetration into 28-mm conventional liners was 132 mum versus 155 mum for highly cross-linked material (P = .017). Ninety percent of the creep had occurred by 2.5 million cycles. (C) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:239 / 243
页数:5
相关论文
共 17 条
[1]
Digas G, 2003, 49 ANN M ORTH RES SO
[2]
Characterization of long-term femoral-head-penetration rates - Association with and prediction of osteolysis [J].
Dowd, JE ;
Sychterz, CJ ;
Young, AM ;
Engh, CA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (08) :1102-1107
[3]
Dowson D, 1988, Eng Med, V17, P111, DOI 10.1243/EMED_JOUR_1988_017_031_02
[4]
A critical assessment of proximal macrotexturing on cemented femoral components [J].
Duffy, GP ;
Muratoglu, OK ;
Biggs, SA ;
Larson, SL ;
Lozynsky, AJ ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :42-48
[5]
The effect of activity levels of total hip arthroplasty patients on socket penetration [J].
Goldsmith, AAJ ;
Dowson, D ;
Wroblewski, BM ;
Siney, PD ;
Fleming, PA ;
Lane, JM .
JOURNAL OF ARTHROPLASTY, 2001, 16 (05) :620-627
[6]
Grobbelaar CJ., 1999, SA BONE JOINT SURG, V9, P140
[7]
In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty [J].
Komistek, RD ;
Dennis, DA ;
Ochoa, JA ;
Haas, BD ;
Hammill, C .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10) :1836-1841
[8]
Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene - A hip-simulator study [J].
McKellop, H ;
Shen, FW ;
Lu, B ;
Campbell, P ;
Salovey, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (12) :1708-1725
[9]
Study of retrieved acetabular sockets made from high-dose, cross-linked polyethylene [J].
Oonishi, H ;
Clarke, IC ;
Masuda, S ;
Amino, H .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :129-133
[10]
Oonishi H, 2001, J BIOMED MATER RES, V58, P167, DOI 10.1002/1097-4636(2001)58:2<167::AID-JBM1003>3.0.CO