A pulse amperometric sensor for the measurement of atmospheric hydrogen peroxide

被引:26
作者
Huang, HL
Dasgupta, PK
Genfa, Z
Wang, J
机构
[1] TEXAS TECH UNIV,DEPT CHEM & BIOCHEM,LUBBOCK,TX 79409
[2] NEW MEXICO STATE UNIV,DEPT CHEM,LAS CRUCES,NM 88003
关键词
D O I
10.1021/ac960143x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gaseous H2O2 is sampled through a Nafion membrane diffusion scrubber while 1 mM HCl is maintained stationary in the scrubber. After a preselected preconcentration time (typically, 5-10 min), a valve is opened to allow the scrubber liquid to now by gravity over an electrochemical H2O2 sensor for a brief period. The miniature flow-over sensor consists of a Pt/Rh wire working electrode and a Pt wire counter electrode wound respectively on separate segments of a Nafion solid polymer electrolyte tubing supported on a Ag/AgCl wire reference electrode. A simple electronic interface and a personal computer are used to control and record the electrochemical measurement. The liquid phase detection limit for this sensor is similar to 30 nM H2O2 in the anodic oxidation mode. For a 9 min gas sample preconcentration period, the LOD (S/N = 3 criterion) is 0.11 ppbv H2O2(g). Ambient H2O2 data obtained with this instrument were in excellent agreement with those obtained by an established fluorometric technique in a blind intercomparison.
引用
收藏
页码:2062 / 2066
页数:5
相关论文
共 28 条
[1]  
ALLEGRINI I, 1993, 41 COMM EUR COMM, P111
[2]   EVALUATION OF VISIBLE-LIGHT PHOTOLYSIS OF OZONE-WATER CLUSTER MOLECULES AS A SOURCE OF ATMOSPHERIC HYDROXYL RADICAL AND HYDROGEN-PEROXIDE [J].
BUCKLEY, PT ;
BIRKS, JW .
ATMOSPHERIC ENVIRONMENT, 1995, 29 (18) :2409-2415
[3]   PEROXIDASE-MODIFIED CARBON-FIBER MICROELECTRODES IN FLOW-THROUGH DETECTION OF HYDROGEN-PEROXIDE AND ORGANIC PEROXIDES [J].
CSOREGI, E ;
GORTON, L ;
MARKOVARGA, G ;
TUDOS, AJ ;
KOK, WT .
ANALYTICAL CHEMISTRY, 1994, 66 (21) :3604-3610
[4]   APPLICATION OF A NESTED LOOP SYSTEM FOR THE FLOW-INJECTION ANALYSIS OF TRACE AQUEOUS PEROXIDES [J].
DASGUPTA, PK ;
HWANG, H .
ANALYTICAL CHEMISTRY, 1985, 57 (06) :1009-1012
[5]   CONTINUOUS LIQUID-PHASE FLUOROMETRY COUPLED TO A DIFFUSION SCRUBBER FOR THE REAL-TIME DETERMINATION OF ATMOSPHERIC FORMALDEHYDE, HYDROGEN-PEROXIDE AND SULFUR-DIOXIDE [J].
DASGUPTA, PK ;
DONG, S ;
HWANG, H ;
YANG, HC ;
GENFA, Z .
ATMOSPHERIC ENVIRONMENT, 1988, 22 (05) :949-963
[6]  
DASGUPTA PK, 1993, ACS SYM SER, V232, P41
[7]   COMPARISON OF COLLECTION DEVICES FOR ATMOSPHERIC PEROXIDES [J].
DESERVES, C ;
ROSS, HB .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1993, 27 (13) :2712-2718
[8]   AVAILABILITY OF H2O2 FOR OXIDATION OF SO2 IN CLOUDS IN THE NORTHEASTERN UNITED-STATES [J].
DUTKIEWICZ, VA ;
BURKHARD, EG ;
HUSAIN, L .
ATMOSPHERIC ENVIRONMENT, 1995, 29 (22) :3281-3292
[9]   AMPEROMETRIC SENSORS FOR PEROXIDE, CHOLINE, AND ACETYLCHOLINE BASED ON ELECTRON-TRANSFER BETWEEN HORSERADISH-PEROXIDASE AND A REDOX POLYMER [J].
GARGUILO, MG ;
HUYNH, N ;
PROCTOR, A ;
MICHAEL, AC .
ANALYTICAL CHEMISTRY, 1993, 65 (05) :523-528
[10]  
GENFA Z, 1991, Analytica Chimica Acta, V243, P207, DOI 10.1016/S0003-2670(00)82562-3