A pulse amperometric sensor for the measurement of atmospheric hydrogen peroxide

被引:26
作者
Huang, HL
Dasgupta, PK
Genfa, Z
Wang, J
机构
[1] TEXAS TECH UNIV,DEPT CHEM & BIOCHEM,LUBBOCK,TX 79409
[2] NEW MEXICO STATE UNIV,DEPT CHEM,LAS CRUCES,NM 88003
关键词
D O I
10.1021/ac960143x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gaseous H2O2 is sampled through a Nafion membrane diffusion scrubber while 1 mM HCl is maintained stationary in the scrubber. After a preselected preconcentration time (typically, 5-10 min), a valve is opened to allow the scrubber liquid to now by gravity over an electrochemical H2O2 sensor for a brief period. The miniature flow-over sensor consists of a Pt/Rh wire working electrode and a Pt wire counter electrode wound respectively on separate segments of a Nafion solid polymer electrolyte tubing supported on a Ag/AgCl wire reference electrode. A simple electronic interface and a personal computer are used to control and record the electrochemical measurement. The liquid phase detection limit for this sensor is similar to 30 nM H2O2 in the anodic oxidation mode. For a 9 min gas sample preconcentration period, the LOD (S/N = 3 criterion) is 0.11 ppbv H2O2(g). Ambient H2O2 data obtained with this instrument were in excellent agreement with those obtained by an established fluorometric technique in a blind intercomparison.
引用
收藏
页码:2062 / 2066
页数:5
相关论文
共 28 条
[12]   ATMOSPHERIC CHEMISTRY OF PEROXIDES - A REVIEW [J].
GUNZ, DW ;
HOFFMANN, MR .
ATMOSPHERIC ENVIRONMENT PART A-GENERAL TOPICS, 1990, 24 (07) :1601-1633
[13]  
HUANG H, 1994, UNPUB
[14]   THERMODYNAMICS OF THE HYDROGEN-PEROXIDE WATER-SYSTEM [J].
HWANG, H ;
DASGUPTA, PK .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1985, 19 (03) :255-258
[15]   FLUOROMETRIC FLOW-INJECTION DETERMINATION OF AQUEOUS PEROXIDES AT NANOMOLAR LEVEL USING MEMBRANE REACTORS [J].
HWANG, H ;
DASGUPTA, PK .
ANALYTICAL CHEMISTRY, 1986, 58 (07) :1521-1524
[16]   TUBULAR MICROPOROUS MEMBRANE ENTRAPPED ENZYME REACTORS FOR FLOW-INJECTION ANALYSIS [J].
HWANG, H ;
DASGUPTA, PK .
ANALYTICAL CHEMISTRY, 1987, 59 (09) :1356-1360
[17]   FLOW-THROUGH AMPEROMETRIC SENSOR FOR HYDROGEN-PEROXIDE MONITORING IN GASEOUS MEDIA [J].
KULYS, J .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 9 (02) :143-147
[18]   AUTOMATED FLUOROMETRIC METHOD FOR HYDROGEN-PEROXIDE IN AIR [J].
LAZRUS, AL ;
KOK, GL ;
LIND, JA ;
GITLIN, SN ;
HEIKES, BG ;
SHETTER, RE .
ANALYTICAL CHEMISTRY, 1986, 58 (03) :594-597
[19]  
SIGG A, 1992, P WORKSH DEV AN TECH
[20]   ELECTROCATALYSIS AND FLOW DETECTION AT A GLASSY-CARBON ELECTRODE MODIFIED WITH A THIN-FILM OF OXYMANGANESE SPECIES [J].
TAHA, Z ;