The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria

被引:50
作者
Almén, A
Tingberg, A
Mattsson, S
Besjakov, J
Kheddache, S
Lanhede, B
Månsson, LG
Zankl, M
机构
[1] Malmo Univ Hosp, Dept Radiat Phys, SE-20502 Malmo, Sweden
[2] Malmo Univ Hosp, Dept Diagnost Radiol, SE-20502 Malmo, Sweden
[3] Sahlgrens Univ Hosp, Dept Diagnost Radiol, SE-41345 Gothenburg, Sweden
[4] Sahlgrens Univ Hosp, Dept Radiat Phys, SE-41345 Gothenburg, Sweden
[5] GSF Natl Res Ctr Environm & Hlth, D-85764 Neuherberg, Germany
关键词
D O I
10.1259/bjr.73.875.11144797
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
In this study we have investigated the image quality of lumbar spine radiographs taken after recording technical and physical parameters. Two technical parameters were altered, tube voltage (70 kV and 90 kV for the anteroposterior (AP) projection and 77 kV and 95 kV for the lateral projection) and sensitivity of the film-screen system (sensitivity class 400 and 600). In total, 85 images were included in the study. Entrance surface dose (ESD) was measured using thermoluminescent dosemeters. The mean value of ESD for the different technique groups varied between 1.9 mGy (90 kV, sensitivity class 400) and 4.6 mGy (70 kV, sensitivity class 400) for the AP projection, and between 6.4 mGy (95 kV, sensitivity class 600) and 20.4 mGy (70 kV, sensitivity class 400) for the lateral projection. Image criteria given in the "European Guidelines on Quality Criteria for Radiographic Images" were used to assess image quality. Two evaluation methods have been employed. A straightforward scoring of fulfilled image criteria, and visual grading analysis using the structures defined in the image criteria. The latter method provided a sharper distinction between groups of images taken using different radiographic techniques. The average number of fulfilled image criteria for the AP projections varied between 0.74 (90 kV, sensitivity class 400) and 0.87 (70 kV, sensitivity class 400). For the lateral projection this number varied between 0.79 (95 kV, sensitivity class 600) and 0.84 (77 kV, sensitivity class 600). This study shows that image criteria are useful tools in clinical studies of image quality.
引用
收藏
页码:1192 / 1199
页数:8
相关论文
共 18 条
[1]
Almen A, 1996, ACTA RADIOL, V37, P337
[2]
Altman DG, 1991, PRACTICAL STAT MED R, P396
[3]
*COMM EUR COMM, 1996, 16635 EUR EN CEC
[4]
Commision of the European Communities, 1996, 16260 EUR EN CEC
[5]
The use of carbon fibre material in radiographic cassettes: Estimation of the dose and contrast advantages [J].
Dance, DR ;
Lester, SA ;
Carlsson, GA ;
Sandborg, M ;
Persliden, J .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832) :383-390
[6]
IN SEARCH OF OPTIMUM CHEST RADIOGRAPHY TECHNIQUES [J].
LEITZ, WK ;
MANSSON, LG ;
HEDBERGVIKSTROM, BRK ;
KHEDDACHE, S .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (784) :314-321
[7]
DOSES TO PATIENTS FROM DIAGNOSTIC-RADIOLOGY IN FRANCE [J].
MACCIA, C ;
BENEDITTINI, M ;
LEFAURE, C ;
FAGNANI, F .
HEALTH PHYSICS, 1988, 54 (04) :397-408
[8]
MCNEIL EA, 1995, RADIAT PROT DOSIM, V57, P437
[10]
PATIENT DOSES AND RISKS FROM DIAGNOSTIC-RADIOLOGY IN NORTHEAST ITALY [J].
PADOVANI, R ;
CONTENTO, G ;
FABRETTO, M ;
MALISAN, MR ;
BARBINA, V ;
GOZZI, G .
BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (710) :155-165