In vivo evaluation of a new sealant material on a rat lung air leak model

被引:41
作者
Kobayashi, H
Sekine, T
Nakamura, T
Shimizu, Y
机构
[1] 3M Hlth Care Ltd, R&D Dept, Sagamihara, Kanagawa 2291185, Japan
[2] Kyoto Univ, Inst Frontier Med Sci, Dept Bioartificial Organs, Sakyo Ku, Kyoto 6068507, Japan
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 2001年 / 58卷 / 06期
关键词
tissue sealant; lung air leak; rat model; albumin-based hydrogel; burst pressure measurement;
D O I
10.1002/jbm.1066
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The ability of an albumin-based hydrogel sealant (ABHS) to prevent air leakage through the suture line after pulmonary surgery was evaluated by comparison with that of a fibrin glue (FG). As an air-leak model, a rat lung was used in which a standard incision was made and the burst pressure for ABHS and FG was measured. The average burst pressures at time 0 for the FG and ABHS groups were 30.8 +/- 15.2 and 77.5 +/- 19.1 mmHg, respectively. At Day 3, the value of ABHS (76.3 +/- 15.8 mmHg) was still significantly higher (P < 0.05) than that of FG (60.0 +/- 21.9 mmHg). At Day 7, no statistical difference was observed between the FG group (71.2 +/- 18.6 mmHg) and the ABHS group (88.8 +/- 11.7 mmHg). Histological examination of the incision at Day 14 revealed that neither sealant was not visible at the incision site, and there was no evidence of adverse tissue reaction. It was concluded that ABHS had good sealing properties and is an alternative to FG for air leakage treatment in pulmonary surgery. (C) 2001 John Wiley & Sons, Inc.
引用
收藏
页码:658 / 665
页数:8
相关论文
共 38 条
[21]  
MATSUDA T, 1989, T AM SOC ART INT ORG, V35, P381
[22]   THE EFFECT OF FIBRIN GLUING TO SEAL BRONCHIAL AND ALVEOLAR LEAKAGES AFTER PULMONARY RESECTIONS AND DECORTICATIONS [J].
MOURITZEN, C ;
DROMER, M ;
KEINECKE, HO .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1993, 7 (02) :75-80
[23]   Sealing effect of rapidly curable gelatin-poly (L-glutamic acid) hydrogel glue on lung air leak [J].
Otani, Y ;
Tabata, Y ;
Ikada, Y .
ANNALS OF THORACIC SURGERY, 1999, 67 (04) :922-926
[24]  
PORTOGHESE M, 1992, PRESSE MED, V21, P1154
[25]   CLINICAL-SIGNIFICANCE OF ANTIBODIES TO BOVINE AND HUMAN THROMBIN AND FACTOR-V AFTER SURGICAL USE OF BOVINE THROMBIN [J].
RAPAPORT, SI ;
ZIVELIN, A ;
MINOW, RA ;
HUNTER, CS ;
DONNELLY, K .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1992, 97 (01) :84-91
[26]  
Seifert J, 1994, J Invest Surg, V7, P167, DOI 10.3109/08941939409015361
[27]  
Sierra D H, 1993, J Biomater Appl, V7, P309, DOI 10.1177/088532829300700402
[28]  
SPOTNITZ WD, 1996, SURG ADHESIVES SEALA, P239
[29]  
TAKAGI M, 1996, J NEW REMEDIES CLIN, V45, P21
[30]   FIBRIN ADHESIVE AND ITS APPLICATION IN THORACIC-SURGERY [J].
THETTER, O .
THORACIC AND CARDIOVASCULAR SURGEON, 1981, 29 (05) :290-292