Effect of modern pacing algorithms on generator longevity: A predictive analysis

被引:21
作者
Gelvan, D
Crystal, E
Dokumaci, B
Goldshmid, Y
Ovsyshcher, IE
机构
[1] Soroka Univ, Med Ctr, Cardiac Res Ctr, Dept Cardiol, Beer Sheva, Israel
[2] Medtronic, Beer Sheva, Israel
[3] Ben Gurion Univ Negev, Fac Hlth Sci, Beer Sheva, Israel
[4] Univ Toronto, Fac Hlth Sci, Sunnybrook & Womens Coll, Hlth Sci Ctr,Div Cardiol, Toronto, ON, Canada
[5] SSK Hosp, Dept Cardiol, Eskisehir, Turkey
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 09期
关键词
pacemaker; battery longevity; capture management; AV interval; sinus preference; threshold; current drain;
D O I
10.1046/j.1460-9592.2003.t01-1-00272.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Pulse generator (PG) longevity is of major importance to the quality of care of pacemaker patients. A series of automatic algorithms affect PG longevity. This study investigated the individual and combined effects of three algorithms incorporated in the Medtronic Kappa 700 pacemaker series: Capture Management periodically measures the stimulation threshold and adjusts the PG output, Sinus Preference allows the sinus rate to prevail in a specified range below the sensor rate, and Search AV allows an extension of the AV interval if spontaneous conduction is observed. The effects of Capture Management, Sinus Preference, and Search AV on device longevity were studied in 21 consecutive patients treated in the VDD and DDDR modes. Patients were followed for 1 year. The data were analyzed using an equation provided by the manufacturer. Capture Management was activated in 20 patients. For 11 PGs at the basic settings, longevity was extended by 5.2%, whereas reprogrammed PGs had no gain. Sinus Preference Was active in four DDDR patients, who gained 12.0 +/- 5.3% atrial sensing from it, with a resultant longevity gain of 1.4 +/- 0.45 months (NS). Search AV was active in 19 patients and 8 responders gained 7.8 +/- 4.4 months PG longevity. The overall longevity in this study was 106.3 +/- 8.4 months with all features as programmed, whereas the longevity without Capture Management and Search AV algorithms would be 98.2 +/- 4.9 months, saving 8.1 +/- 5.8 months (range 0-18) of battery life. Thus, two algorithms: Capture Management and Search AV, have clinical relevance in the extension of PG longevity.
引用
收藏
页码:1796 / 1802
页数:7
相关论文
共 18 条
[1]  
Antretter H, 1996, ACTA MED AUST, V23, P159
[2]  
Boriani G, 2000, PACE, V23, P1783
[3]  
Crossley GH, 1996, CIRCULATION, V94, P245
[4]   Clinical use of low output settings in 1.2-mm2 steroid eluting electrodes:: Three years of experience [J].
Danilovic, D ;
Ohm, OJ ;
Breivik, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (12) :2606-2615
[5]   Steroid eluting high impedance pacing leads decrease short and long term current drain: Results from a multicenter clinical trial [J].
Ellenbogen, KA ;
Wood, MA ;
Gilligan, DM ;
Zmijewski, M ;
Mans, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (01) :39-48
[6]   Longevity of dual chamber pacemakers:: Device and patient related determinants [J].
Kindermann, M ;
Schwaab, B ;
Berg, M ;
Fröhlig, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (05) :810-815
[7]   STIMULATION CHARACTERISTICS OF A STEROID-ELUTING ELECTRODE COMPARED WITH 3 CONVENTIONAL ELECTRODES [J].
KLEIN, HH ;
STEINBERGER, J ;
KNAKE, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (02) :134-137
[8]  
LINDEMANS FW, 2002, COMMUNICATION 0910
[9]  
MENEZES AS, 1999, HEARTWEB, V4
[10]   High impedance low energy pacing leads: Long-term results with a very small surface area steroid-eluting lead compared to three conventional electrodes [J].
Moracchini, PV ;
Cornacchia, D ;
Bernasconi, M ;
Tesorieri, MC ;
Fabbri, M ;
Marzegalli, M ;
Baraldi, P ;
Corte, T ;
Giuliani, M ;
Marotta, T ;
De Seta, F .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (02) :326-334