Dialyzer Reuse with Peracetic Acid Does Not Impact Patient Mortality

被引:16
作者
Bond, T. Christopher [1 ]
Nissenson, Allen R. [1 ]
Krishnan, Mahesh [1 ]
Wilson, Steven M. [1 ]
Mayne, Tracy [1 ]
机构
[1] DaVita Inc, El Segundo, CA 90245 USA
来源
CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY | 2011年 / 6卷 / 06期
关键词
HEMODIALYSIS-PATIENTS; SURVIVAL; ASSOCIATION;
D O I
10.2215/CJN.10391110
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background and objectives Numerous studies have shown the overall benefits of dialysis filter reuse, including superior biocompatibility and decreased nonbiodegradable medical waste generation, without increased risk of mortality. A recent study reported that dialyzer reprocessing was associated with decreased patient survival; however, it did not control for sources of potential confounding. We sought to determine the effect of dialyzer reprocessing with peracetic acid on patient mortality using contemporary outcomes data and rigorous analytical techniques. Design, setting, participants, & measurements We conducted a series of analyses of hemodialysis patients examining the effects of reuse on mortality using three techniques to control for potential confounding: instrumental variables, propensity-score matching, and time-dependent survival analysis. Results In the instrumental variables analysis, patients at high reuse centers had 16.2 versus 15.9 deaths/100 patient-years in nonreuse centers. In the propensity-score matched analysis, patients with reuse had a lower death rate per 100 patient-years than those without reuse (15.2 versus 15.5). The risk ratios for the time-dependent survival analyses were 0.993 (per percent of sessions with reuse) and 0.995 (per unit of last reuse), respectively. Over the study period, 13.8 million dialyzers were saved, representing 10,000 metric tons of medical waste. Conclusions Despite the large sample size, powered to detect miniscule effects, neither the instrumental variables nor propensity-matched analyses were statistically significant. The time-dependent survival analysis showed a protective effect of reuse. These data are consistent with the preponderance of evidence showing reuse limits medical waste generation without negatively affecting clinical outcomes. Clin J Am Soc Nephrol 6: 1368-1374, 2011. doi: 10.2215/CJN.10391110
引用
收藏
页码:1368 / 1374
页数:7
相关论文
共 14 条
[11]  
Parsons L.S., 2001, P 26 ANN SAS USERS G
[12]   Mortality risk by hemodialyzer reuse practice and dialyzer membrane characteristics: Results from the USRDS dialysis morbidity and mortality study [J].
Port, FK ;
Wolfe, RA ;
Hulbert-Shearon, TE ;
Daugirdas, JT ;
Agodoa, LYC ;
Jones, C ;
Orzol, SM ;
Held, PJ .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2001, 37 (02) :276-286
[13]   Dialyzer reuse - Part II: Advantages and disadvantages [J].
Twardowski, ZJ .
SEMINARS IN DIALYSIS, 2006, 19 (03) :217-226
[14]  
Yan Lily, 2007, Nephrol News Issues, V21, P58