Study of the retention properties of warfarin enantiomers on a β-cyclodextrin polymeric support

被引:20
作者
Guillaume, M [1 ]
Jaulmes, A [1 ]
Sébille, B [1 ]
Thuaud, N [1 ]
Vidal-Madjar, C [1 ]
机构
[1] Univ Paris 12, CNRS, Lab Rech Polymeres, UMR C7581, F-94320 Thiais, France
来源
JOURNAL OF CHROMATOGRAPHY B | 2001年 / 753卷 / 01期
关键词
enantiomer separation; beta-cyclodextrin polymeric support; retention properties; warfarin;
D O I
10.1016/S0378-4347(00)00371-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
High-performance liquid chromatography was used to study the retention properties of (R)- and (S)-warfarins on a silica support coated with a beta -cyclodextrin polymer. The influence of the methanol content of the acetate buffer eluent was investigated at pH 4. The measure of the variations of retention time with temperature enables one to determine the enthalpy and the entropy of adsorption. The plot of the two thermodynamic functions shows a minimum around 30% (v/v) methanol. At low methanol contents, the decrease of the hydrophobic interactions with increasing methanol content explains the decrease of the enthalpic and entropic terms. Above 40% (v/v) methanol, the decrease of the adsorption enthalpy absolute value is due to the solvation by the organic component. From the analysis of peak shape in mass-overload conditions, the column capacity toward each enantiomer was determined. A lower capacity was found toward (S)-warfarin, the more retained enantiomer. Peak shape analysis in mass-overload conditions was used to determine the adsorption isotherm. A Langmuir-type adsorption isotherm accounts well for the experimental data. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 22 条
[1]   THERMODYNAMIC STUDY OF THE MARKED DIFFERENCES BETWEEN ACETONITRILE WATER AND METHANOL WATER MOBILE-PHASE SYSTEMS IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
ALVAREZZEPEDA, A ;
BARMAN, BN ;
MARTIRE, DE .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1978-1984
[2]   SEPARATION OF METALLOCENE ENANTIOMERS BY LIQUID-CHROMATOGRAPHY - CHIRAL RECOGNITION VIA CYCLODEXTRIN BONDED PHASES [J].
ARMSTRONG, DW ;
DEMOND, W ;
CZECH, BP .
ANALYTICAL CHEMISTRY, 1985, 57 (02) :481-484
[3]   SEPARATION OF DRUG STEREOISOMERS BY THE FORMATION OF BETA-CYCLODEXTRIN INCLUSION COMPLEXES [J].
ARMSTRONG, DW ;
WARD, TJ ;
ARMSTRONG, RD ;
BEESLEY, TE .
SCIENCE, 1986, 232 (4754) :1132-1135
[4]   (R)-NAPHTHYLETHYLCARBAMATE-SUBSTITUTED AND (S)-NAPHTHYLETHYLCARBAMATE-SUBSTITUTED BETA-CYCLO-DEXTRIN BONDED STATIONARY PHASES FOR THE REVERSED-PHASE LIQUID-CHROMATOGRAPHIC SEPARATION OF ENANTIOMERS [J].
ARMSTRONG, DW ;
CHANG, CD ;
LEE, SH .
JOURNAL OF CHROMATOGRAPHY, 1991, 539 (01) :83-90
[5]   THEORETICAL CONSIDERATIONS CONCERNING THE SEPARATION OF ENANTIOMERIC SOLUTES BY LIQUID-CHROMATOGRAPHY [J].
BOEHM, RE ;
MARTIRE, DE ;
ARMSTRONG, DW .
ANALYTICAL CHEMISTRY, 1988, 60 (06) :522-528
[6]   RETENTION OF SOME MONOAMINE-OXIDASE INHIBITORY DRUGS ON A BETA-CYCLODEXTRIN POLYMER-COATED SILICA COLUMN [J].
CSERHATI, T ;
FORGACS, E .
JOURNAL OF CHROMATOGRAPHY A, 1994, 660 (1-2) :313-318
[7]   RETENTION BEHAVIOR OF BARBITURIC-ACID DERIVATIVES ON A BETA-CYCLODEXTRIN POLYMER-COATED SILICA COLUMN [J].
FORGACS, E ;
CSERHATI, T .
JOURNAL OF CHROMATOGRAPHY A, 1994, 668 (02) :395-402
[8]   RETENTION BEHAVIOR AND CHIRAL RECOGNITION MECHANISM OF SEVERAL CYCLODEXTRIN-BONDED STATIONARY PHASES FOR DANSYL AMINO-ACIDS [J].
FUJIMURA, K ;
SUZUKI, S ;
HAYASHI, K ;
MASUDA, S .
ANALYTICAL CHEMISTRY, 1990, 62 (20) :2198-2205
[9]   Study of a polymer immobilized on silica gel as stationary phase for chiral liquid chromatography [J].
Guillaume, M ;
Sebille, B .
EUROPEAN POLYMER JOURNAL, 1996, 32 (01) :19-26
[10]  
Han SM, 1989, CHIRAL SEPARATIONS H, P208