A finite element model of a microwave catheter for cardiac ablation

被引:14
作者
Kaouk, Z [1 ]
Khebir, A [1 ]
Savard, P [1 ]
机构
[1] UNIV MONTREAL,MONTREAL,PQ,CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1109/22.539943
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
To investigate the delivery of microwave energy by a catheter located inside the heart for the purpose of ablating small abnormal regions producing cardiac arrhythmias, a numerical model was developed, This model is based oil the finite element method and call solver both the electromagnetic field and the temperature distribution resulting from the radiated power for axisymmetrical geometries. The antenna, which is fed by a coaxial cable with a 2.4 mm diameter, is constituted by a monopole which is terminated by a metallic cylindrical cap, The heart model can be either homogeneous or constituted of coaxial cylindrical shells with different electrical and thermal conductivities representing the intracavitary blood masses, the heart, and the torso. Experimental measurements obtained in an homogeneous tissue equivalent medium, such as the reflection coefficient of the antenna al different frequencies and for different monopole lengths, the radial and axial steady-state temperature profiles, and the time course of the temperature rise, were all ill close agreement with the values computed with the model, Accurate modeling is a useful prerequisite for the design of antennas, and these results confirm the validity of the catheter-heart model for the investigation and the development of microwave catheters.
引用
收藏
页码:1848 / 1854
页数:7
相关论文
共 23 条
[11]   EPICARDIAL AND ENDOCARDIAL MAPPING OF VENTRICULAR-TACHYCARDIA IN PATIENTS WITH MYOCARDIAL-INFARCTION - IS THE ORIGIN OF THE TACHYCARDIA ALWAYS SUBENDOCARDIALLY LOCALIZED [J].
KALTENBRUNNER, W ;
CARDINAL, R ;
DUBUC, M ;
SHENASA, M ;
NADEAU, R ;
TREMBLAY, G ;
VERMEULEN, M ;
SAVARD, P ;
PAGE, PL .
CIRCULATION, 1991, 84 (03) :1058-1071
[12]  
KAOUK Z, 1996, MED BIOL ENG COMPUT, V34, P1
[13]  
KHEBIR A, 1995, IEEE MTT S ORL
[14]  
LABONTE S, 1995, IEEE MTT-S, P303, DOI 10.1109/MWSYM.1995.406046
[15]   A COMPUTER-SIMULATION OF RADIOFREQUENCY ABLATION OF THE ENDOCARDIUM [J].
LABONTE, S .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (09) :883-890
[16]  
LEE JF, 1993, IEEE T MICROW THEORY, V41, P1981
[17]  
NATH S, 1992, J AM COLL CARDIOL, V19
[18]  
ROSENBAUM RM, 1993, IEEE MTT S ATL
[19]   NORTH-AMERICAN-SOCIETY-OF-PACING-AND-ELECTROPHYSIOLOGY (NASPE) SURVEY ON RADIOFREQUENCY CATHETER ABLATION - IMPLICATIONS FOR CLINICIAN, 3RD PARTY INSURERS, AND GOVERNMENT REGULATORY AGENCIES [J].
SCHEINMAN, MM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (12) :2228-2231
[20]  
SILVESTER P., 1983, FINITE ELEMENTS ELEC