THA loading arising from increased femoral anteversion and offset may lead to critical cement stresses

被引:74
作者
Kleemann, RU [1 ]
Heller, MO [1 ]
Stoeckle, U [1 ]
Taylor, WR [1 ]
Duda, GN [1 ]
机构
[1] Humboldt Univ, Charite, D-13353 Berlin, Germany
关键词
anteversion; joint geometry; muscle forces; total hip arthroplasty; finite element analysis;
D O I
10.1016/S0736-0266(03)00040-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aseptic loosening of artificial hip joints is believed to be influenced by the design and orientation of the implant. It is hypothesised that variations in implant anteversion and offset lead to changes in the loading of the proximal femur, causing critical conditions to both the bone and cement. The goal of this study was therefore to analyse the role of these parameters on loading, bone strains and cement stresses in total hip arthroplasty (THA). A validated musculo-skeletal model was used for the analysis of muscle and joint contact forces during walking and stair climbing. Two different anteversion angles (4degrees vs. 24degrees) and prostheses offsets (standard vs. long) were analysed. The loads for each case were applied to a cemented THA finite element model. Generally, stair climbing caused higher bone strains and cement stresses (max. +25%) than walking. Variations in anteversion and offset caused changes in the loading environment, bone strain distribution and cement stresses. Compared to the standard THA configuration, cement stresses were raised by increasing anteversion (max. +52%), offset (max. +5%) and their combination (max. +67%). Femoral anteversion, offset and their combination may therefore lead to an increased risk of implant loosening. Analyses of implant survival should consider this as a limiting factor in THA longevity. In clinical practice, implant orientation, especially in regard to pre- and post-operative anteversion, should be considered to be more critical. (C) 2003 Orthopaedic Research Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:767 / 774
页数:8
相关论文
共 42 条
[1]   The pathology of total joint arthroplasty - II. Mechanisms of implant failure [J].
Bauer, TW ;
Schils, J .
SKELETAL RADIOLOGY, 1999, 28 (09) :483-497
[2]   The pathology of total joint arthroplasty I. Mechanisms of implant fixation [J].
Bauer, TW ;
Schils, J .
SKELETAL RADIOLOGY, 1999, 28 (08) :423-432
[3]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[4]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]   IS STAIRCASE WALKING A RISK FOR THE FIXATION OF HIP IMPLANTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :535-553
[6]   Hybrid total hip arthroplasty with a precoated offset stem - Four to nine-year results [J].
Cannestra, VP ;
Berger, RA ;
Quigley, LR ;
Jacobs, JJ ;
Rosenberg, AG ;
Galante, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09) :1291-1299
[7]  
CHANG PB, 1998, CLIN ORTHOP RELAT R, V355, P57
[8]   ROLE OF LOADS AND PROSTHESIS MATERIAL PROPERTIES ON THE MECHANICS OF THE PROXIMAL FEMUR AFTER TOTAL HIP-ARTHROPLASTY [J].
CHEAL, EJ ;
SPECTOR, M ;
HAYES, WC .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1992, 10 (03) :405-422
[9]  
Davey J R, 1993, J Arthroplasty, V8, P23
[10]   SUPERIOR DISPLACEMENT OF THE HIP IN TOTAL JOINT REPLACEMENT - EFFECTS OF PROSTHETIC NECK LENGTH, NECK-STEM ANGLE, AND ANTEVERSION ANGLE ON THE MOMENT-GENERATING CAPACITY OF THE MUSCLES [J].
DELP, SL ;
KOMATTU, AV ;
WIXSON, RL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1994, 12 (06) :860-870