Packed capillary reversed-phase liquid chromatography with high-performance electrospray ionization Fourier transform ion cyclotron resonance mass spectrometry for proteomics

被引:208
作者
Shen, YF [1 ]
Zhao, R [1 ]
Belov, ME [1 ]
Conrads, TP [1 ]
Anderson, GA [1 ]
Tang, KQ [1 ]
Pasa-Tolic, L [1 ]
Veenstra, TD [1 ]
Lipton, MS [1 ]
Udseth, HR [1 ]
Smith, RD [1 ]
机构
[1] Pacific NW Natl Lab, Environm Mol Sci Lab, Richland, WA 99352 USA
关键词
D O I
10.1021/ac0011336
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, high-efficiency packed capillary reversed-phase liquid chromatography (RPLC) coupled on-line with high-performance Fourier transform ion cyclotron resonance (FTICR) mass spectrometry has been investigated for the characterization of complex cellular proteolytic digests. Long capillary columns (80-cm) packed with small (3-mum) C18 bonded particles provided a total peak capacity of similar to 1000 for cellular proteolytic polypeptides when interfaced with an ESI-FTICR mass spectrometer under composition gradient conditions at a pressure of 10 000 psi, Large quantities of cellular proteolytic digests (e.g,, 500 mug) could be loaded onto packed capillaries of 150-mum inner diameter without a significant loss of separation efficiency. Precolumns with suitable inner diameters were found useful for improving the elution reproducibility without a significant loss of separation quality. Porous particle packed capillaries were found to provide better results than those containing nonporous particles because of their higher sample capacity. Two-dimensional analyses from the combination of packed capillary RPLC with high-resolution FTICR yield a combined capacity for separations of >1 million polypeptide components and simultaneously provided information for the identification of the separated components based upon the accurate mass tag concept previously described.
引用
收藏
页码:1766 / 1775
页数:10
相关论文
共 51 条
[21]  
Görg A, 2000, ELECTROPHORESIS, V21, P1037, DOI 10.1002/(SICI)1522-2683(20000401)21:6<1037::AID-ELPS1037>3.0.CO
[22]  
2-V
[23]   Evaluation of two-dimensional gel electrophoresis-based proteome analysis technology [J].
Gygi, SP ;
Corthals, GL ;
Zhang, Y ;
Rochon, Y ;
Aebersold, R .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2000, 97 (17) :9390-9395
[24]  
Gygi SP, 1999, ELECTROPHORESIS, V20, P310, DOI 10.1002/(SICI)1522-2683(19990201)20:2<310::AID-ELPS310>3.0.CO
[25]  
2-M
[26]   Quantitative analysis of complex protein mixtures using isotope-coded affinity tags [J].
Gygi, SP ;
Rist, B ;
Gerber, SA ;
Turecek, F ;
Gelb, MH ;
Aebersold, R .
NATURE BIOTECHNOLOGY, 1999, 17 (10) :994-999
[27]  
Harry JL, 2000, ELECTROPHORESIS, V21, P1071, DOI 10.1002/(SICI)1522-2683(20000401)21:6<1071::AID-ELPS1071>3.3.CO
[28]  
2-D
[29]   IDENTIFYING PROTEINS FROM 2-DIMENSIONAL GELS BY MOLECULAR MASS SEARCHING OF PEPTIDE-FRAGMENTS IN PROTEIN-SEQUENCE DATABASES [J].
HENZEL, WJ ;
BILLECI, TM ;
STULTS, JT ;
WONG, SC ;
GRIMLEY, C ;
WATANABE, C .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1993, 90 (11) :5011-5015
[30]   HIGH-FLOW ION SPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
HOPFGARTNER, G ;
WACHS, T ;
BEAN, K ;
HENION, J .
ANALYTICAL CHEMISTRY, 1993, 65 (04) :439-446