RTD analysis of a novel Taylor-Couette flow device for blood detoxification

被引:26
作者
Ameer, GA
Grovender, EA
Obradovic, B
Cooney, CL
Langer, R [1 ]
机构
[1] MIT, Dept Chem Engn, Cambridge, MA 02139 USA
[2] MIT, Ctr Biotechnol Proc Engn, Cambridge, MA 02139 USA
关键词
D O I
10.1002/aic.690450320
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Heparin is an anticoagulant used in extracorporeal procedures such as hemodialysis and open heart surgery. Unfortunately: heparin may induce potentially fatal complications in patients at high risk of bleeding. The use of art immobilized heparinase I reactor makes heparin therapy safer; but the design of a safe and efficient reactor for medical use had been a significant problem. A novel reactor, based on simultaneous separation-reaction and Taylor-Couette flow, was designed and successfully tested in vitro with human blood and ex vivo in sleep. The objective of this study was to understand the flow dynamics in the reactor in order to predict and optimize heparin neutralization. Residence-time distribution studies were performed and a mathematical model was developed The model was able to predict experimental conversions within a mean relative error of 5.5%. Bypass flow through the reactive section was also predicted.
引用
收藏
页码:633 / 638
页数:6
相关论文
共 20 条
[1]   Ex vivo evaluation of a Taylor-Couette flow, immobilized heparinase I device for clinical application [J].
Ameer, GA ;
Barabino, G ;
Sasisekharan, R ;
Harmon, W ;
Cooney, CL ;
Langer, R .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1999, 96 (05) :2350-2355
[2]  
Ameer GA, 1999, BIOTECHNOL BIOENG, V62, P602, DOI 10.1002/(SICI)1097-0290(19990305)62:5<602::AID-BIT12>3.0.CO
[3]  
2-M
[4]  
AMEER GA, 1999, BIOTECHNOL BIOENG, V63
[5]  
BEAUDOIN G, 1987, LIFE SUPPORT SYST, V5, P273
[6]   Expression in Escherichia coli, purification and characterization of heparinase I from Flavobacterium heparinum [J].
Ernst, S ;
Venkataraman, G ;
Winkler, S ;
Godavarti, R ;
Langer, R ;
Cooney, CL ;
Sasisekharan, R .
BIOCHEMICAL JOURNAL, 1996, 315 :589-597
[7]  
FISCHEL RJ, 1988, T AM SOC ART INT ORG, V34, P375
[8]  
HEUSER G, 1980, BIORHEOLOGY, V17, P17
[9]   QUANTITATIVE DESCRIPTION OF ULTRAFILTRATION IN A ROTATING FILTRATION DEVICE [J].
HOLESCHOVSKY, UB ;
COONEY, CL .
AICHE JOURNAL, 1991, 37 (08) :1219-1226
[10]  
HORROW JC, 1985, ANESTH ANALG, V64, P348