Three-dimensional reconstruction allows accurate quantification and length measurements of coronary artery stenoses

被引:23
作者
Rittger, Harald [1 ]
Schertel, Birgit [1 ]
Schmidt, Martin [1 ]
Justiz, Joern [2 ]
Brachmann, Johannes [1 ]
Sinha, Anil-Martin [1 ]
机构
[1] Klinikum Coburg, Med Klin 2, D-96450 Coburg, Germany
[2] Fachhsch Bern, Fachbereich Med Tech, Bern, Switzerland
关键词
Coronary artery stenosis; quantitative coronary angiography; 3D-reconstruction; length measurements; percutaneous coronary intervention;
D O I
10.4244/EIJV5I1A20
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: The aim of this prospective study was to evaluate the feasibility and accuracy of a recently developed 3D system (CardiOp-B; Paieon Medical Ltd., Israel) as compared to a validated quantitative coronary angiography (QCA) system (Siemens Quantcor, Siemens Medical Solutions). Methods and results: In patients scheduled for heart catheterisation, minimal lumen diameter (MLD) and diameter-derived percent stenosis (DPS) were obtained for CAS (>50%) using both QCA and the 3D-system. To estimate stenosis length, a non-inflated balloon was inserted into the stenosis and the distance between balloon markers was measured using both methods and then compared to the known distance between the markers. In 61 patients 79 lesions were analysed. MLD measurements showed a good agreement between QCA and 3D with a mean difference of 0.08 +/- 0.035 mm. Reference diameter was 2.61 +/- 0.67 for 3D and 2.42 +/- 0.61 mm for QCA and 54.79 +/- 9.20% vs. 58.75 +/- 8.15% for the %-stenosis range, respectively. The mean true balloon length was 12.8 mm +/- 3.8 mm. Lengths determined by the 3D system were 13.0 +/- 4.0 mm and 11.3 +/- 3.8 mm by QCA, respectively. Conclusions: Evaluation of CAS using the novel 3D system was feasible and showed equivalent results to validated QCA measurements. Length measurements seemed to be more accurate by the 3D system as compared to QCA. Therefore, this 3D-system can be used to guide decisions in interventional cardiology.
引用
收藏
页码:127 / 132
页数:6
相关论文
共 20 条
[1]   Quantitative analysis of reconstructed 3-D coronary arterial tree and intracoronary devices [J].
Chen, SYJ ;
Carroll, JD ;
Messenger, JC .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (07) :724-740
[2]  
DVIR D, 2007, EUROINTERVENTION, V2, P141
[3]  
Dvir Danny, 2005, Int J Cardiovasc Intervent, V7, P141
[4]   QUANTITATIVE CORONARY ANGIOGRAPHY (QCA) IN INTERVENTIONAL CARDIOLOGY - CLINICAL-APPLICATION OF QCA MEASUREMENTS [J].
FOLEY, DP ;
ESCANED, J ;
STRAUSS, BH ;
DIMARIO, C ;
HAASE, J ;
KEANE, D ;
HERMANS, WRM ;
RENSING, BJ ;
DEFEYTER, PJ ;
SERRUYS, PW .
PROGRESS IN CARDIOVASCULAR DISEASES, 1994, 36 (05) :363-384
[5]   In vivo validation of intravascular ultrasound length measurements using a motorized transducer pullback system [J].
Fuessl, RT ;
Mintz, GS ;
Pichard, AD ;
Kent, KM ;
Satler, LF ;
Popma, JJ ;
Leon, MB .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (12) :1115-&
[6]   Utility of three-dimensional reconstruction of coronary angiography to guide percutaneous coronary intervention [J].
Gollapudi, Raghava R. ;
Valencia, Rafael ;
Lee, Steve S. ;
Wong, Garrett B. ;
Teirstein, Paul S. ;
Price, Matthew J. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (04) :479-482
[7]  
GottsaunerWolf M, 1996, EUR HEART J, V17, P1167
[8]   Clinical assessment of a new real time 3D quantitative coronary angiography system: Evaluation in stented vessel segments [J].
Gradaus, R ;
Mathies, K ;
Breithardt, G ;
Boecker, D .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (01) :44-49
[9]   Angiographic views used for percutaneous coronary interventions: A three-dimensional analysis of physician-determined vs. computer-generated views [J].
Green, NE ;
Chen, SYJ ;
Hansgen, AR ;
Messenger, JC ;
Groves, BM ;
Carroll, JD .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (04) :451-459
[10]   CAAS-II - A 2ND GENERATION SYSTEM FOR OFF-LINE AND ONLINE QUANTITATIVE CORONARY ANGIOGRAPHY [J].
GRONENSCHILD, E ;
JANSSEN, J ;
TIJDENS, F .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 33 (01) :61-75