Counting efficiency of the API Aerosizer

被引:20
作者
Thornburg, J [1 ]
Cooper, SJ [1 ]
Leith, D [1 ]
机构
[1] Univ N Carolina, Dept Environm Sci & Engn, Chapel Hill, NC 27599 USA
关键词
Number:; R010H03293; Acronym:; NIOSH; Sponsor: National Institute for Occupational Safety and Health; -; ED; Sponsor: U.S. Department of Education;
D O I
10.1016/S0021-8502(98)00073-1
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The Aerosizer (Amherst Process Instruments, Inc. Hadley MA) is a time-of-Right instrument frequently used to measure the size distribution of an aerosol. However, if the Aerosizer's counting efficiency, defined as the number of particles counted divided by the total number entering the instrument, is not 100% or varies with particle size, the resulting size distribution will be inaccurate. Experiments were conducted to determine the effect of particle diameter, particle concentration, photomultiplier tube (PMT) voltage, and model type on the Aerosizer's counting efficiency. To calculate counting efficiency, the number of particles between 0.3 and 10 mu m recorded by the Aerosizer was divided by the number of particles of the same size collected on each stage of a cascade impactor. Particle diameter, aerosol concentration, Aerosizer model, PMT voltage, and the diameter interaction terms influenced counting efficiency. Counting efficiencies were less than 1 % for particles smaller than 0.45 mu m, and more than 100% for particles larger than 7 mu m. Increasing the PMT voltage increased the counting efficiency for the smaller particles, but also created false, larger particles. Counting efficiency decreased as count rate increased for count rates greater than 20,000 particles per second. The Aerosizer LD counted particles more efficiently than the Aerosizer Mach 2 because of improved laser and optics systems. Four regression models that relate counting efficiency to the salient operating parameters were developed, one for each combination of Aerosizer model and photomultiplier tube voltage studied. (C) 1999 Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:479 / 488
页数:10
相关论文
共 14 条
[1]  
ANDERSEN G, 1991, SERIES 210 AMBIENT C
[2]  
*API INC, 1992, AER MACH 2 US MAN VE
[3]  
*API INC, 1995, AER LD US MAN VERS 7
[4]  
BARON PA, 1996, AM ASS AER RES 15 AN
[5]  
BARON PA, 1993, AEROSOL MEASUREMENT, P403
[6]   CALIBRATION AND PERFORMANCE OF AN API AEROSIZER [J].
CHENG, YS ;
BARR, EB ;
MARSHALL, IA ;
MITCHELL, JP .
JOURNAL OF AEROSOL SCIENCE, 1993, 24 (04) :501-514
[7]   PROPERTIES OF CONTINUUM SOURCE PARTICLE BEAMS .2. BEAMS GENERATED IN CAPILLARY EXPANSIONS [J].
CHENG, YS ;
DAHNEKE, BE .
JOURNAL OF AEROSOL SCIENCE, 1979, 10 (04) :363-368
[8]   AEROSOL BEAM SPECTROMETRY [J].
DAHNEKE, B .
NATURE-PHYSICAL SCIENCE, 1973, 244 (134) :54-55
[9]   PROPERTIES OF CONTINUUM SOURCE PARTICLE BEAMS .1. CALCULATION METHODS AND RESULTS [J].
DAHNEKE, BE ;
CHENG, YS .
JOURNAL OF AEROSOL SCIENCE, 1979, 10 (03) :257-274
[10]   INLET EFFICIENCY STUDY FOR THE TSI AERODYNAMIC PARTICLE SIZER [J].
KINNEY, PD ;
PUI, DYH .
PARTICLE & PARTICLE SYSTEMS CHARACTERIZATION, 1995, 12 (04) :188-193