Enzymatically modified ion-selective electrodes for flow injection analysis

被引:19
作者
Koncki, R [1 ]
Walcerz, I [1 ]
Leszczynska, E [1 ]
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
关键词
flow injection analysis; ion-selective electrodes; enzyme immobilization; urea; penicillins;
D O I
10.1016/S0731-7085(98)00230-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The application of potentiometric biosensors based on membrane ion-selective electrodes for flow injection analysis (FIA) is presented. The biosensors have been obtained by covalent binding of enzyme molecules directly to surface of ion-selective membranes. The biosensor/FIA systems enable the determination of urea and penicillin in the millimolar ranges of concentration. About 30 samples per hour can be analysed in the performed FIA systems. The operational stability of the bioanalytical systems exceeds 1 month. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:633 / 638
页数:6
相关论文
共 9 条
[1]  
Cammann K., 1979, WORKING ION SELECTIV
[2]  
KAUFFMANN JM, 1991, BIOSENSOR PRINCIPLES, P63
[3]   Penicillin enzyme biosensors based on pH membrane electrodes [J].
Koncki, R ;
Leszczynska, E .
ANALYTICA CHIMICA ACTA, 1996, 321 (01) :27-34
[4]  
KONCKI R, 1991, CHEM ANAL-WARSAW, V36, P423
[6]  
RILEY M, 1979, GAS SENSING PROBES I
[7]  
Valcarcel M., 1994, Flow-through, Bio)Chemical Sensors
[8]  
VALCAREL M, 1987, FLOW INJECTION ANAL
[9]  
Walcerz I, 1996, ANAL LETT, V29, P1939