Improvement and extension of the black-body-based spectral irradiance scale

被引:28
作者
Sperfeld, P [1 ]
Metzdorf, J [1 ]
Yousef, SG [1 ]
Stock, KD [1 ]
Möller, W [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
D O I
10.1088/0026-1394/35/4/9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Using a new, high-temperature black-body radiator developed by the All-Russian Institute for Optophysical Measurements (VNIIOFI), Moscow, and applying newly developed methods and equipment, the black-body-based spectral irradiance scale of the Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, has been extended and its uncertainties have been reduced compared with previous results. The pyrolytic-graphite Planckian radiator BB3200pg, which can be operated at temperatures up to over 3300 K, is presented and characterized along with modifications to optimize uniformity and stability. The determination of the radiometric black-body temperatures is already well established by measuring the weighted irradiance using filter detectors without imaging systems. This method is compared and verified with results based on self-consistent spectral irradiance measurements using fast-scanning spectroradiometers. In addition, a laser-based method for the determination of the radiometric temperature is presented and compared with conventional procedures of radiation-temperature measurement. The implementation of this additional method into the radiometric chain of the PTB is illustrated. Improvements in the uncertainty budget of the spectral irradiance scale are discussed.
引用
收藏
页码:267 / 271
页数:5
相关论文
共 12 条
[1]  
[Anonymous], P SPIE
[2]   International comparison of radiation-temperature measurements with filtered detectors over the temperature range 1380 K to 3100 K [J].
Harrison, NJ ;
Fox, NP ;
Sperfeld, P ;
Metzdorf, J ;
Khlevnoy, BB ;
Stolyarevskaya, RI ;
Khromchenko, VB ;
Mekhontsev, SN ;
Shapoval, VI ;
Zelener, MF ;
Sapritsky, VI .
METROLOGIA, 1998, 35 (04) :283-288
[3]   NETWORK AND TRACEABILITY OF THE RADIOMETRIC AND PHOTOMETRIC STANDARDS AT THE PTB [J].
METZDORF, J .
METROLOGIA, 1993, 30 (04) :403-408
[4]  
METZDORF J, 1993, BFSISH16393, P15
[5]  
METZDORF J, 1987, IMEKO TC SERIES, V17, P89
[6]  
METZDORF J, 1996, ADV METROLOGY ITS RO, P297
[7]   SPECTRORADIOMETRIC DETERMINATION OF THE FREEZING TEMPERATURE OF GOLD [J].
MIELENZ, KD ;
SAUNDERS, RD ;
SHUMAKER, JB .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1990, 95 (01) :49-67
[8]  
Möller W, 1998, METROLOGIA, V35, P261, DOI 10.1088/0026-1394/35/4/8
[9]   Precision blackbody sources for radiometric standards [J].
Sapritsky, VI ;
Khlevnoy, BB ;
Khromchenko, VB ;
Lisiansky, BE ;
Mekhontsev, SN ;
Melenevsky, UA ;
Morozova, SP ;
Prokhorov, AV ;
Samoilov, LN ;
Shapoval, VI ;
Sudarev, KA ;
Zelener, MF .
APPLIED OPTICS, 1997, 36 (22) :5403-5408
[10]   Spectral-irradiance scale based on radiometric black-body temperature measurements [J].
Sperfeld, P ;
Raatz, KH ;
Nawo, B ;
Moller, W ;
Metzdorf, J .
METROLOGIA, 1996, 32 (06) :435-439