Simultaneous anodic stripping voltammetric determination of lead and cadmium with a carbon paste electrode modified by tributyl phosphate

被引:13
作者
Huang, SS [1 ]
Cheng, YD [1 ]
Li, BF [1 ]
Liu, GD [1 ]
机构
[1] Hunan Univ, Dept Chem & Chem Engn, Changsha 410082, Peoples R China
关键词
voltammetry; CPE; lead; cadmium;
D O I
10.1007/BF01254596
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A chemically modified carbon paste electrode was prepared by incorporating quantities of tributyl phosphate (TBP) directly into the paste mixture, In the presence of 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP), trace lead and cadmium in acetic acid-acetate medium can be preconcentrated on the electrode. The method showed good linearly for 1.0 x 10(-8) - 8.0 x 10(-7) mol.L-1 lead(II) and for 5.0 x 10(-8) - 2.0 x 10(-6) mol.L-1 cadmium(II). Many common metal ions have no or little effect on the determination of lead(II) and cadmium(II). The electrode was verified by the determination of trace lead(II) and cadmium(II) in waste water and human hair samples.