Modified perturbation method for transverse electromagnetic (TEM) coil tuning and evaluation

被引:11
作者
Avdievich, NI
Krymov, VN
Hetherington, HP
机构
[1] Yeshiva Univ Albert Einstein Coll Med, Dept Radiol, Bronx, NY 10461 USA
[2] Gruss Magnet Resonance Res Ctr, Bronx, NY USA
[3] Albert Einstein Coll Med, Dept Physiol & Biophys, Bronx, NY 10467 USA
关键词
TEM head coil bench tuning; MRI; perturbation method; angular distribution of current; RF magnetic and electric field mapping;
D O I
10.1002/mrm.10539
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A modified perturbation method for tuning and testing volume head coils was developed. The common perturbation method, utilizing the change in resonance frequency of a resonator in response to the presence of a small dielectric or magnetic probe, was modified to modulate the frequency shift due to rotation of a probe. This modification enabled the RIF magnetic and electric fields as well as the angular distribution of current in the longitudinal elements of the coil to be mapped. The latter serves as a quick test of magnetic field homogeneity by comparing the measured distribution with the sinusoidal function required for the field to be homogenous. The method was experimentally applied to tune and test a transverse electromagnetic (TEM) head coil. Magn Reson Med 50:13-18, 2003. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:13 / 18
页数:6
相关论文
共 19 条
[1]  
Barfuss H, 1990, NMR Biomed, V3, P31, DOI 10.1002/nbm.1940030106
[2]   SMALL NMR PROBE FOR USE IN INHOMOGENEOUS FIELDS [J].
BREUER, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1965, 36 (11) :1666-&
[3]   Practical aspects of birdcage coils [J].
Doty, FD ;
Entzminger, G ;
Hauck, CD ;
Staab, JP .
JOURNAL OF MAGNETIC RESONANCE, 1999, 138 (01) :144-154
[4]   RADIATIVE LOSSES OF A BIRDCAGE RESONATOR [J].
HARPEN, MD .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (05) :713-716
[5]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[6]   EVALUATION OF CEREBRAL GRAY AND WHITE-MATTER METABOLITE DIFFERENCES BY SPECTROSCOPIC IMAGING AT 4.1T [J].
HETHERINGTON, HP ;
MASON, GF ;
PAN, JW ;
PONDER, SL ;
VAUGHAN, JT ;
TWIEG, DB ;
POHOST, GM .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (05) :565-571
[7]   Resonant modes of the birdcage coil [J].
Leifer, MC .
JOURNAL OF MAGNETIC RESONANCE, 1997, 124 (01) :51-60
[8]   FIELD STRENGTH MEASUREMENTS IN RESONANT CAVITIES [J].
MAIER, LC ;
SLATER, JC .
JOURNAL OF APPLIED PHYSICS, 1952, 23 (01) :68-77
[9]  
MASCART E, 1886, LECONS ELECTRICITE M, V1, P357
[10]   Quantitative spectroscopic imaging of the human brain [J].
Pan, JW ;
Twieg, DB ;
Hetherington, HP .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (03) :363-369