A multiwavelength frequency-domain near-infrared cerebral oximeter

被引:45
作者
Kurth, CD
Thayer, WS
机构
[1] Childrens Hosp Philadelphia, Dept Anesthesiol & Crit Care Med, Philadelphia, PA 19104 USA
[2] Univ Penn, Sch Med, Dept Anesthesia Physiol & Pediat, Philadelphia, PA 19104 USA
[3] NIM Inc, Philadelphia, PA 19104 USA
关键词
D O I
10.1088/0031-9155/44/3/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study tests a multiwavelength frequency-domain near-infrared oximeter(fdNIRS) in an in vitro model of the human brain. The model is a solid plastic structure containing a vascular network perfused with blood in which haemoglobin oxygen saturation (SO2) was measured by co-oximetry, providing a standard for comparison. Plastic shells of varying thickness (0.5-2 cm), with a vascular network of their own and encircling the brain model, were also added to simulate extracranial tissues of the infant, child and adult. The fdNIRS oximeter utilizes frequency-domain technology to monitor phaseshifts at 754 nm, 785 nm and 816 nm relative to a 780 nm reference to derive SO2 through photon transport and Beer-Lambert equations. We found a linear relationship between fdNIRS SO2 and co-oximetry SO2 with excellent correlation (r(2) greater than or equal to 0.95) that fitted the line of identity in all experiments (n = 7). The bias of fdNIRS oximetry was -2% and the precision was 6%. Blood temperature and fdNIRS source-detector distance did not affect fdNIRS oximetry. Low haemoglobin concentration (6 g dl(-1)) altered the fdNIRS versus co-oximetry line slope and intercept, producing a 15% error at the extremes of SO2. The infant- and child-like shells overlying the brain model did not alter fdNIRS oximetry, whereas the adult-like shell yielded an error as high as 32%. In conclusion, fdNIRS accurately measures SO2 in an in vitro brain model, although low haemoglobin concentration and extracranial tissue of adult thickness influence accuracy.
引用
收藏
页码:727 / 740
页数:14
相关论文
共 29 条
[1]   THE THEORETICAL BASIS FOR THE DETERMINATION OF OPTICAL PATHLENGTHS IN TISSUE - TEMPORAL AND FREQUENCY-ANALYSIS [J].
ARRIDGE, SR ;
COPE, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (07) :1531-1560
[2]   TRANSCRANIAL OPTICAL-PATH LENGTH IN INFANTS BY NEAR-INFRARED PHASE-SHIFT SPECTROSCOPY [J].
BENARON, DA ;
KURTH, CD ;
STEVEN, JM ;
DELIVORIAPAPADOPOULOS, M ;
CHANCE, B .
JOURNAL OF CLINICAL MONITORING, 1995, 11 (02) :109-117
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Cerebral oxygenation measured by near-infrared spectroscopy: Comparison with jugular bulb oximetry [J].
Daubeney, PEF ;
Pilkington, SN ;
Janke, E ;
Charlton, GA ;
Smith, DC ;
Webber, SA .
ANNALS OF THORACIC SURGERY, 1996, 61 (03) :930-934
[5]   CEREBRAL AND MUSCLE OXYGEN-SATURATION MEASUREMENT BY FREQUENCY-DOMAIN NEAR-INFRA-RED SPECTROMETER [J].
DEBLASI, RA ;
FANTINI, S ;
FRANCESCHINI, MA ;
FERRARI, M ;
GRATTON, E .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1995, 33 (02) :228-230
[6]   ESTIMATION OF OPTICAL PATHLENGTH THROUGH TISSUE FROM DIRECT TIME OF FLIGHT MEASUREMENT [J].
DELPY, DT ;
COPE, M ;
VANDERZEE, P ;
ARRIDGE, S ;
WRAY, S ;
WYATT, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (12) :1433-1442
[7]   Measurement of cranial optical path length as a function of age using phase resolved near infrared spectroscopy [J].
Duncan, A ;
Meek, JH ;
Clemence, M ;
Elwell, CE ;
Fallon, P ;
Tyszczuk, L ;
Cope, M ;
Delpy, DT .
PEDIATRIC RESEARCH, 1996, 39 (05) :889-894
[8]   FREQUENCY-DOMAIN MULTICHANNEL OPTICAL-DETECTOR FOR NONINVASIVE TISSUE SPECTROSCOPY AND OXIMETRY [J].
FANTINI, S ;
FRANCESCHINI, MA ;
MAIER, JS ;
WALKER, SA ;
BARBIERI, B ;
GRATTON, E .
OPTICAL ENGINEERING, 1995, 34 (01) :32-42
[9]   MEASUREMENT OF THE OPTICAL-PROPERTIES OF THE SKULL IN THE WAVELENGTH RANGE 650-950 NM [J].
FIRBANK, M ;
HIRAOKA, M ;
ESSENPREIS, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (04) :503-510
[10]   NEAR-INFRARED SPECTROSCOPY IN ADULTS - EFFECTS OF EXTRACRANIAL ISCHEMIA AND INTRACRANIAL HYPOXIA ON ESTIMATION OF CEREBRAL OXYGENATION [J].
GERMON, TJ ;
KANE, NM ;
MANARA, AR ;
NELSON, RJ .
BRITISH JOURNAL OF ANAESTHESIA, 1994, 73 (04) :503-506