A mouse model to evaluate the interface between skin and a percutaneous device

被引:36
作者
Isenhath, S. N.
Fukano, Y.
Usui, M. L.
Underwood, R. A.
Irvin, C. A.
Marshall, A. J.
Hauch, K. D.
Ratner, B. D.
Fleckman, P.
Olerud, J. E. [1 ]
机构
[1] Univ Washington, Dept Dermatol Med, Seattle, WA 98195 USA
[2] Univ Washington, Dept BioEngn, Seattle, WA 98195 USA
关键词
wound healing; poly(HEMA); implant; biointegration; biomaterial;
D O I
10.1002/jbm.a.31391
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Percutaneous medical devices are integral in the management and treatment of disease. The space created between the skin and the device becomes a haven for bacterial invasion and biofilm formation and results in infection. We hypothesize that sealing this space via integration of the skin into the device will create a barrier against bacterial invasion. The purpose of this study was to develop an animal model in which the interaction between skin and biomaterials can be evaluated. Porous poly(2-hydroxyethyl methacrylate) [poly(HEMA)l rods were implanted for 7 days in the dorsal skin of C57 BL/6 mice. The porous poly(HEMA) rods were surface-modified with carbonyldiimidazole (CDI) or CDI plus laminin 5; unmodified rods served as control. Implant sites were sealed with 2-octyl cyanoacrylate; corn pads and adhesive dressings were tested for stabilization of implants. All rods remained intact for the duration of the study. There was histological evidence of both epidermal and dermal integration into all poly(HEMA) rods regardless of treatment. This in vivo model permits examination of the implant/skin interface and will be useful for future studies designed to facilitate skin cell attachment where percutaneous devices penetrate the skin. (c) 2007 Wiley Periodicals, Inc.
引用
收藏
页码:915 / 922
页数:8
相关论文
共 38 条
[1]  
Bhende Shubhangi, 2002, Surg Infect (Larchmt), V3, P251, DOI 10.1089/109629602761624216
[2]   Closure of long surgical incisions with a new formulation of 2-octylcyanoacrylate tissue adhesive versus commercially available methods [J].
Blondeel, PNV ;
Murphy, JW ;
Debrosse, D ;
Nix, JC ;
Puls, LE ;
Theodore, N ;
Coulthard, P .
AMERICAN JOURNAL OF SURGERY, 2004, 188 (03) :307-313
[3]  
Bruns TB, 2000, AM FAM PHYSICIAN, V61, P1383
[4]   A LIGHT AND ELECTRON-MICROSCOPIC STUDY OF THE EFFECTS OF SURFACE-TOPOGRAPHY ON THE BEHAVIOR OF CELLS ATTACHED TO TITANIUM-COATED PERCUTANEOUS IMPLANTS [J].
CHEHROUDI, B ;
GOULD, TRL ;
BRUNETTE, DM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (03) :387-405
[5]   Subcutaneous microfabricated surfaces inhibit epithelial recession and promote long-term survival of percutaneous implants [J].
Chehroudi, B ;
Brunette, DM .
BIOMATERIALS, 2002, 23 (01) :229-237
[6]   The promise of novel technology for the prevention of intravascular device-related bloodstream infection. I. Pathogenesis and short-term devices [J].
Crnich, CJ ;
Maki, DG .
CLINICAL INFECTIOUS DISEASES, 2002, 34 (09) :1232-1242
[7]  
Deporter D A, 1985, Prog Clin Biol Res, V187, P353
[8]   Assessment of anticollagenase treatments after insertion of a keratoprosthetic material in the rabbit cornea [J].
Fitton, JH ;
Ziegelaar, BW ;
Hicks, CR ;
Clayton, AB ;
Crawford, GJ ;
Constable, IJ ;
Chirila, TV .
CORNEA, 1998, 17 (01) :108-114
[9]   Characterization of an in vitro model for evaluating the interface between skin and percutaneous biomaterials [J].
Fukano, Yuko ;
Knowles, Negar G. ;
Usui, Marcia L. ;
Underwood, Robert A. ;
Hauch, Kip D. ;
Marshall, Andrew J. ;
Ratner, Buddy D. ;
Giachelli, Cecilia ;
Carter, William G. ;
Fleckman, Philip ;
Olerud, John E. .
WOUND REPAIR AND REGENERATION, 2006, 14 (04) :484-491
[10]   Updated feasibility trial experience with the viaderm percutaneous access device [J].
Gesler, W ;
Smith, R ;
DeDecker, PG ;
Bernstam, L ;
Snyder, R ;
Freed, PS ;
Kantrowitz, A .
ASAIO JOURNAL, 2004, 50 (04) :349-353