Comparison of the NovaSom QSG™, a new sleep apnea home-diagnostic system, and polysomnography

被引:61
作者
Reichert, JA
Bloch, DA
Cundiff, E
Votteri, BA
机构
[1] Sequoia Hosp, Sleep Disorders Ctr, Redwood City, CA 94062 USA
[2] Stanford Univ, Sch Med, Dept Hlth Res & Policy, Div Biostat, Stanford, CA 94305 USA
关键词
home monitoring; three nights; unattended; self-administered; obstructive sleep apnea; diagnosis; polysomnography;
D O I
10.1016/S1389-9457(02)00234-4
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Obstructive sleep apnea (OSA) is a serious, common, and underdiagnosed disorder that challenges health care resources. While polysomnography (PSG) represents the standard diagnostic test for OSA, portable devices provide an alternative diagnostic tool when issues of cost, time, geographic availability, or other constraints pose impediments to in-lab testing. This study compares the NovaSom QSG(TM), a new sleep apnea home diagnostic system, to PSG both in the laboratory and in the home. Methods: Fifty-one consecutive adults referred to the sleep lab for suspicion of OSA underwent one night of in-lab, simultaneous recording of PSG and NovaSom QSG in addition to using the NovaSom QSG at home for three nights. Two separate comparisons were made using the apnea-hypopnea index (AHI): in-lab PSG to in-lab NovaSom QSG and in-lab PSG to home NovaSom QSG. Results: Using a clinical cut-off of AHI = 15, the sensitivity and specificity of the in-lab NovaSom QSG vs. PSG were 95% and 91%, respectively. For home NovaSom QSG vs. in-lab PSG, the sensitivity was 91% and specificity was 83%. The intra-class correlation coefficient for the agreement between three separate nights of NovaSom QSG home data was 0.88. Conclusions: In a patient population suspected of having OSA, the NovaSom QSG demonstrated acceptable sensitivity and specificity both in the lab and self-administered in the home, when compared to PSG. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:213 / 218
页数:6
相关论文
共 30 条
[1]  
Amer Acad Sleep Med, 2000, SLEEP, V23, P1017
[2]   The association of sleep-disordered breathing and sleep symptoms with quality of life in the sleep heart health study [J].
Baldwin, CM ;
Griffith, KA ;
Nieto, FJ ;
O'Connor, GT ;
Walsleben, JA ;
Redline, S .
SLEEP, 2001, 24 (01) :96-105
[3]   Sleep apnea in pat ischemic attack and stroke: A prospective study of 59 patients [J].
Bassetti, C ;
Aldrich, MS ;
Chervin, RD ;
Quint, D .
NEUROLOGY, 1996, 47 (05) :1167-1173
[4]   MEASUREMENT ERROR IN VISUALLY SCORED ELECTROPHYSIOLOGICAL DATA - RESPIRATION DURING SLEEP [J].
BLIWISE, D ;
BLIWISE, NG ;
KRAEMER, HC ;
DEMENT, W .
JOURNAL OF NEUROSCIENCE METHODS, 1984, 12 (01) :49-56
[5]  
Chesson AL, 1997, SLEEP, V20, P406
[6]   Clinical validation of the Bedbugg™ in detection of obstructive sleep apnea [J].
Claman, D ;
Murr, A ;
Trotter, K .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2001, 125 (03) :227-230
[7]   Scoring variability between polysomnography technologists in different sleep laboratories [J].
Collop, Nancy A. .
SLEEP MEDICINE, 2002, 3 (01) :43-47
[8]   SLEEP-APNEA AND HYPERTENSION - A POPULATION-BASED STUDY [J].
HLA, KM ;
YOUNG, TB ;
BIDWELL, T ;
PALTA, MM ;
SKATRUD, B ;
DEMPSEY, J .
ANNALS OF INTERNAL MEDICINE, 1994, 120 (05) :382-388
[9]  
Horstmann S, 2000, SLEEP, V23, P383
[10]   The medical cost of undiagnosed sleep apnea [J].
Kapur, V ;
Blough, DK ;
Sandblom, RE ;
Hert, R ;
de Maine, JB ;
Sullivan, SD ;
Psaty, BM .
SLEEP, 1999, 22 (06) :749-755