Clinical validation of the Bedbugg™ in detection of obstructive sleep apnea

被引:22
作者
Claman, D
Murr, A
Trotter, K
机构
[1] Univ Calif San Francisco, Sleep Disorders Ctr, San Francisco, CA 94115 USA
[2] Univ Calif San Francisco, Dept Otolaryngol, San Francisco, CA 94115 USA
关键词
D O I
10.1067/mhn.2001.118126
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
OBJECTIVE: To validate the accuracy of the Bedbugg(TM), a new home monitoring device for diagnosis of obstructive sleep apnea. STUDY DESIGN AND SETTING: Simultaneous sleep monitoring was performed by formal polysomnography and by Bedbugg. Monitoring was performed in a university sleep center in 42 subjects who had previously been scheduled for polysomnography. RESULTS: The correlation for the apnea-hypopnea index (AHI) between polysomnography and Bedbugg was r = 0.96. The sensitivity of Bedbugg for detecting an AHI > 15 was 85.7%. The specificity of Bedbugg for detecting an AHI < 15 was 95.2%. CONCLUSION: The Bedbugg device provides an accurate assessment of the apnea-hypopnea index. SIGNIFICANCE: Accurate home monitoring for sleep apnea may provide access to care for a higher proportion of undiagnosed sleep apnea patients.
引用
收藏
页码:227 / 230
页数:4
相关论文
共 11 条
[1]   PORTABLE RECORDING IN THE ASSESSMENT OF OBSTRUCTIVE SLEEP-APNEA [J].
FERBER, R ;
MILLMAN, R ;
COPPOLA, M ;
FLEETHAM, J ;
MURRAY, CF ;
IBER, C ;
MCCALL, V ;
NINOMURCIA, G ;
PRESSMAN, M ;
SANDERS, M ;
STROHL, K ;
VOTTERI, B ;
WILLIAMS, A .
SLEEP, 1994, 17 (04) :378-392
[2]   THE SLEEP HYPOPNEA SYNDROME [J].
GOULD, GA ;
WHYTE, KF ;
RHIND, GB ;
AIRLIE, MAA ;
CATTERALL, JR ;
SHAPIRO, CM ;
DOUGLAS, NJ .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 137 (04) :895-898
[3]   A CAUSE OF EXCESSIVE DAYTIME SLEEPINESS - THE UPPER AIRWAY-RESISTANCE SYNDROME [J].
GUILLEMINAULT, C ;
STOOHS, R ;
CLERK, A ;
CETEL, M ;
MAISTROS, P .
CHEST, 1993, 104 (03) :781-787
[4]   The role of the primary care physician in recognizing obstructive sleep apnea [J].
Kramer, NR ;
Cook, TE ;
Carlisle, CC ;
Corwin, RW ;
Millman, RP .
ARCHIVES OF INTERNAL MEDICINE, 1999, 159 (09) :965-968
[5]   Indications for positive airway pressure treatment of adult obstructive sleep apnea patients - A consensus statement [J].
Loube, DI ;
Gay, PC ;
Strohl, KP ;
Pack, AI ;
White, DP ;
Collop, NA .
CHEST, 1999, 115 (03) :863-866
[6]   VALIDATION OF A PORTABLE SLEEP-APNEA MONITORING DEVICE [J].
MAN, GCW ;
KANG, BV .
CHEST, 1995, 108 (02) :388-393
[7]   Portable computerized polysomnography in attended and unattended settings [J].
Mykytyn, IJ ;
Sajkov, D ;
Neill, AM ;
McEvoy, RD .
CHEST, 1999, 115 (01) :114-122
[8]   Hypopnea, a floating metric: Implications for prevalence, morbidity estimates, and case finding [J].
Redline, S ;
Sanders, M .
SLEEP, 1997, 20 (12) :1209-1217
[9]   ASSESSMENT OF ACCURACY AND ANALYSIS TIME OF A NOVEL DEVICE TO MONITOR SLEEP AND BREATHING IN THE HOME [J].
WHITE, DP ;
GIBB, TJ ;
WALL, JM ;
WESTBROOK, PR .
SLEEP, 1995, 18 (02) :115-126
[10]   THE OCCURRENCE OF SLEEP-DISORDERED BREATHING AMONG MIDDLE-AGED ADULTS [J].
YOUNG, T ;
PALTA, M ;
DEMPSEY, J ;
SKATRUD, J ;
WEBER, S ;
BADR, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1993, 328 (17) :1230-1235