Selenium radiography versus storage phosphor and conventional radiography in the detection of simulated chest lesions

被引:52
作者
SchaeferProkop, CM [1 ]
Prokop, M [1 ]
Schmidt, A [1 ]
Neitzel, U [1 ]
Galanski, M [1 ]
机构
[1] PHILIPS MED SYSTS,HAMBURG,GERMANY
关键词
phantoms; radiography; digital; technology; receiver operating characteristic curve (ROC); thorax;
D O I
10.1148/radiology.201.1.8816519
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare selenium detectors with three conventional and digital detector systems for the detection of simulated pulmonary lesions. MATERIALS AND METHODS: Templates containing nodules, linear structures, and micronodular opacities were superimposed over an anthropomorphic chest phantom. The authors compared lesion detection with use of storage phosphor radiography (250 speed), selenium radiography (250 speed) with an antiscatter grid, selenium radiography (450 speed) without an antiscatter grid, an asymmetric screen-film system (400 speed), and a conventional screenfilm system (250 speed). Detection performance of 10 radiologists was compared by using a multireader-multicase receiver operating characteristic analysis of variance. RESULTS: For the detection of nodules, no statistically significant differences between imaging modes were seen. For the detection of micronodules and linear lesions, both selenium techniques were superior to all other modes (P < .05). In addition, the asymmetric screen-film radiographs were inferior (P < .05) to the conventional screen-film radiographs and to storage phosphor radiographs for the detection of micronodules. CONCLUSION: The selenium detector improves detection of simulated fine linear and low-contrast micronodular details and appears to be superior to other detector systems for chest radiography.
引用
收藏
页码:45 / 50
页数:6
相关论文
共 18 条
[1]   TECHNICAL EVALUATION OF A DIGITAL CHEST RADIOGRAPHY SYSTEM THAT USES A SELENIUM DETECTOR [J].
CHOTAS, HG ;
FLOYD, CE ;
RAVIN, CE .
RADIOLOGY, 1995, 195 (01) :264-270
[2]   THRESHOLD PERCEPTION PERFORMANCE WITH COMPUTED AND SCREEN-FILM RADIOGRAPHY - IMPLICATIONS FOR CHEST RADIOGRAPHY [J].
DOBBINS, JT ;
RICE, JJ ;
BEAM, CA ;
RAVIN, CE .
RADIOLOGY, 1992, 183 (01) :179-187
[3]   MAXIMUM-LIKELIHOOD ESTIMATION OF PARAMETERS OF SIGNAL-DETECTION THEORY AND DETERMINATION OF CONFIDENCE INTERVALS - RATING-METHOD DATA [J].
DORFMAN, DD ;
ALF, E .
JOURNAL OF MATHEMATICAL PSYCHOLOGY, 1969, 6 (03) :487-&
[4]   RECEIVER OPERATING CHARACTERISTIC RATING ANALYSIS - GENERALIZATION TO THE POPULATION OF READERS AND PATIENTS WITH THE JACKKNIFE METHOD [J].
DORFMAN, DD ;
BERBAUM, KS ;
METZ, CE .
INVESTIGATIVE RADIOLOGY, 1992, 27 (09) :723-731
[5]  
FLOYD CE, 1994, P SOC PHOTO-OPT INS, V2163, P110
[6]  
HEESEWIJK JPM, 1995, EUR RADIOL, V5, P83
[7]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .4. EFFECT OF UNSHARP MASKING ON THE DETECTABILITY OF SIMPLE PATTERNS [J].
LOO, LND ;
DOI, K ;
METZ, CE .
MEDICAL PHYSICS, 1985, 12 (02) :209-214
[8]  
METZ C, 1987, SEMIN NUCL MED, V8, P283
[9]  
Metz CE, 1990, 1990 JOINT STAT M AM
[10]   IMAGE QUALITY OF A DIGITAL CHEST RADIOGRAPHY SYSTEM BASED ON A SELENIUM DETECTOR [J].
NEITZEL, U ;
MAACK, I ;
GUNTHERKOHFAHL, S .
MEDICAL PHYSICS, 1994, 21 (04) :509-516