A comparison of temperature measurements using three ear thermometers

被引:4
作者
Weiss, ME
Sitzer, V
Clarke, M
Haley, K
Richards, M
Sanchez, A
Gocka, I
机构
[1] Marquette Univ, Coll Nursing, Milwaukee, WI 53201 USA
[2] Sharp Mem Hosp, San Diego, CA USA
关键词
D O I
10.1016/S0897-1897(98)80274-7
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Following anecdotal and research-based reports of inaccuracies encountered with the use of ear thermometers in patient care settings, this two-phase study was designed to compare the accuracy of three ear thermometers when used in a multioperator environment. The within-subject variation (limits of agreement) in ear temperature measurements obtained with the three ear thermometers in a multioperator condition by using an oral temperature reference standard ranged from +/-1.25 degrees F to 1.85 degrees F. In multioperator, multisubject patient care environments using a pulmonary artery catheter core temperature as the reference standard. the limits of agreement for ear temperature measurements obtained with the three different ear thermometers ranged from +/-2.19 degrees F to 2.85 degrees F. These results suggest that there is substantial variation in ear measurements and raise questions about the use of ear thermometers when there are multiple personnel taking temperatures, as occurs in hospital-based clinical practice environments. Handedness of the operator, position of the patient, and the ear used for measurement did not produce clinically significant variability in ear temperature measurements. Copyright (C) 1998 by W.B. Saunders Company.
引用
收藏
页码:158 / 166
页数:9
相关论文
共 19 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   COMPARISON OF A TYMPANIC THERMOMETER TO RECTAL AND ORAL THERMOMETERS IN A PEDIATRIC EMERGENCY DEPARTMENT [J].
CHAMBERLAIN, JM ;
GRANDNER, J ;
RUBINOFF, JL ;
KLEIN, BL ;
WAISMAN, Y ;
HUEY, M .
CLINICAL PEDIATRICS, 1991, 30 (04) :24-29
[3]  
Erickson R S, 1994, Am J Crit Care, V3, P40
[4]   COMPARISON OF EAR-BASED, BLADDER, ORAL, AND AXILLARY METHODS FOR CORE TEMPERATURE-MEASUREMENT [J].
ERICKSON, RS ;
KIRKLIN, SK .
CRITICAL CARE MEDICINE, 1993, 21 (10) :1528-1534
[5]  
ERICKSON RS, 1994, HEART LUNG, V23, P181
[6]  
ERICKSON RS, 1991, NURS RES, V40, P90
[7]  
FREED GL, 1992, PEDIATRICS, V89, P384
[8]   COMPARISON OF RECTAL, AXILLARY, AND TYMPANIC MEMBRANE TEMPERATURES IN INFANTS AND YOUNG-CHILDREN [J].
MUMA, BK ;
TRELOAR, DJ ;
WURMLINGER, K ;
PETERSON, E ;
VITAE, A .
ANNALS OF EMERGENCY MEDICINE, 1991, 20 (01) :41-44
[9]  
NIELD M, 1993, NURS RES, V42, P294
[10]   INFRARED EAR THERMOMETRY [J].
NOBEL, JJ .