Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter

被引:522
作者
Siewerdsen, JH [1 ]
Jaffray, DA [1 ]
机构
[1] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
关键词
flat-panel imager; cone-beam computed tomography; x-ray scatter; scatter-to-primary ratio; artifacts; contrast; noise; contrast-to-noise ratio;
D O I
10.1118/1.1339879
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A system for cone-beam computed tomography (CBCT) based on a flat-panel imager (FPI) is used to examine the magnitude and effects of x-ray scatter in FPI-CBCT volume reconstructions. The system is being developed for application in image-guided therapies and has previously demonstrated spatial resolution and soft-tissue visibility comparable or superior to a conventional CT scanner under conditions of low x-ray scatter. For larger objects consistent with imaging of human anatomy (e.g., the pelvis) and for increased cone angle (i.e., larger volumetric reconstructions), however, the effects of x-ray scatter become significant. The magnitude of x-ray scatter with which the FPI-CBCT system must contend is quantified in terms of the scatter-to-primary energy fluence ratio (SPR) and scatter intensity profiles in the detector plane, each measured as a function of object size and cone angle. For large objects and cone angles (e.g., a pelvis imaged with a cone angle of 6 degrees), SPR in excess of 100% is observed. Associated with such levels of x-ray scatter are cup and streak artifacts as well as reduced accuracy in reconstruction values, quantified herein across a range of SPR consistent with the clinical setting. The effect of x-ray scatter on the contrast, noise, and contrast-to-noise ratio (CNR) in FPI-CBCT reconstructions was measured as a function of SPR and compared to predictions of a simple analytical model. The results quantify the degree to which elevated SPR degrades the CNR. For example, FPI-CBCT images of a breast-equivalent insert in water were degraded in CNR by nearly a factor of 2 for SPR ranging from similar to2% to 120%. The analytical model for CNR provides a quantitative understanding of the relationship between CNR, dose, and spatial resolution and allows knowledgeable selection of the acquisition and reconstruction parameters that, for a given SPR, are required to restore the CNR to values achieved under conditions of low x-ray scatter. For example, for SPR=100%, the CNR in FPI-CBCT images can be fully restored by: (1) increasing the dose by a factor of 4 (at full spatial resolution); (2) increasing dose and slice thickness by a factor of 2; or (3) increasing slice thickness by a factor of 4 (with no increase in dose). Other reconstruction parameters, such as transaxial resolution length and reconstruction filter, can be similarly adjusted to achieve CNR equal to that obtained in the scatter-free case. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:220 / 231
页数:12
相关论文
共 37 条
[1]  
*AM I PHYS, 1991, 31 AAPM AM I PHYS
[2]  
BARRETT H H, 1976, Computers in Biology and Medicine, V6, P307, DOI 10.1016/0010-4825(76)90068-8
[3]   AN ANALYTICAL MODEL OF THE SCATTERED RADIATION DISTRIBUTION IN DIAGNOSTIC-RADIOLOGY [J].
BOONE, JM ;
SEIBERT, JA .
MEDICAL PHYSICS, 1988, 15 (05) :721-725
[4]   PHYSICAL CHARACTERISTICS OF SCATTERED RADIATION IN DIAGNOSTIC-RADIOLOGY - MONTE-CARLO SIMULATION STUDIES [J].
CHAN, HP ;
DOI, K .
MEDICAL PHYSICS, 1985, 12 (02) :152-165
[5]   CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS [J].
CHO, PS ;
JOHNSON, RH ;
GRIFFIN, TW .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1863-1883
[6]  
CLACK R, 1994, P SOC PHOTO-OPT INS, V2299, P230, DOI 10.1117/12.179253
[7]   CONTRAST-DETAIL-DOSE ANALYSIS OF 6 DIFFERENT COMPUTED TOMOGRAPHIC SCANNERS [J].
COHEN, G .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (02) :197-203
[8]   Effect of scatter radiation on image noise in cone beam CT [J].
Endo, M ;
Tsunoo, T ;
Nakamori, N .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :514-521
[9]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[10]   COMPTON SCATTER EFFECTS IN CT RECONSTRUCTIONS [J].
GLOVER, GH .
MEDICAL PHYSICS, 1982, 9 (06) :860-867