A study of dipole localization accuracy for MEG and EEC using a human skull phantom

被引:273
作者
Leahy, RM [1 ]
Mosher, JC
Spencer, ME
Huang, MX
Lewine, JD
机构
[1] Univ So Calif, Inst Signal & Image Proc, Los Angeles, CA 90089 USA
[2] Los Alamos Natl Lab, Biophys Grp, Los Alamos, NM 87545 USA
[3] Signal Proc Solut Inc, Redondo Beach, CA 90277 USA
[4] Univ New Mexico, Dept Radiol, Albuquerque, NM 87131 USA
[5] Univ Utah, Dept Radiol, Salt Lake City, UT 84132 USA
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1998年 / 107卷 / 02期
关键词
magnetoencephalography (MEG); electroencephalography (EEG); dipole localization; phantom; boundary element methods; locally fitted spheres;
D O I
10.1016/S0013-4694(98)00057-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: To investigate the accuracy of forward and inverse techniques for EEG and MEG dipole localization. Design and Methods: A human skull phantom was constructed with brain, skull and scalp layers and realistic relative conductivities. Thirty two independent current dipoles were distributed within the 'brain' region and EEG and MEG data collected separately for each dipole. The true dipole locations and orientations and the morphology of the brain, skull and scalp layers were extracted from X-ray CT data. The location of each dipole was estimated from the EEG and MEG data using the R-MUSIC inverse method and forward models based on spherical and realistic head geometries. Additional computer simulations were performed to investigate the factors affecting localization accuracy. Results: Localization errors using the relatively simpler locally fitted sphere approach are only slightly greater than those using a BEM approach. The average localization error over the 32 dipoles was 7-8 mm for EEG and 3 mm for MEG, Conclusion: The superior performance of MEG over EEG appears to be because the latter is more sensitive to errors in the forward model arising from simplifying assumptions concerning the conductivity of the skull, scalp and brain. (C) 1998 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:159 / 173
页数:15
相关论文
共 22 条
  • [1] BAILLET S, 1998, IN PRESS HUM BRAIN M
  • [2] MAGNETIC LOCALIZATION OF A DIPOLAR CURRENT SOURCE IMPLANTED IN A SPHERE AND A HUMAN CRANIUM
    BARTH, DS
    SUTHERLING, W
    BROFFMAN, J
    BEATTY, J
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1986, 63 (03): : 260 - 273
  • [3] ECCENTRIC DIPOLE IN A SPHERICAL MEDIUM - GENERALIZED EXPRESSION FOR SURFACE POTENTIALS
    BRODY, DA
    TERRY, FH
    IDEKER, RE
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (02) : 141 - 143
  • [4] MEG VERSUS EEG LOCALIZATION TEST USING IMPLANTED SOURCES IN THE HUMAN BRAIN
    COHEN, D
    CUFFIN, BN
    YUNOKUCHI, K
    MANIEWSKI, R
    PURCELL, C
    COSGROVE, GR
    IVES, J
    KENNEDY, JG
    SCHOMER, DL
    [J]. ANNALS OF NEUROLOGY, 1990, 28 (06) : 811 - 817
  • [5] A COMPLETE LINEAR DISCRETIZATION FOR CALCULATING THE MAGNETIC-FIELD USING THE BOUNDARY-ELEMENT METHOD
    FERGUSON, AS
    ZHANG, X
    STROINK, G
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (05) : 455 - 460
  • [6] Golub GH, 2013, Matrix Computations, V4
  • [7] GREENBLATT RE, 1994, BRAIN TOPOGR, V6, P331
  • [8] REALISTIC CONDUCTIVITY GEOMETRY MODEL OF THE HUMAN HEAD FOR INTERPRETATION OF NEUROMAGNETIC DATA
    HAMALAINEN, MS
    SARVAS, J
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (02) : 165 - 171
  • [9] ILMONIEMI RJ, 1985, BIOMAGNETISM APPL TH, P278
  • [10] Lewine JD, 1995, STUD APPL ELECTROMAG, V7, P368