Inductively-coupled power and data link for neural prostheses using a Class-E oscillator and FSK modulation

被引:47
作者
Troyk, PR [1 ]
DeMichele, GA [1 ]
机构
[1] IIT, Chicago, IL 60616 USA
来源
PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH | 2003年 / 25卷
关键词
transcutaneous; link; Class-E; modulation; neural prosthesis;
D O I
10.1109/IEMBS.2003.1280869
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Despite the success of implantable batteries as commonly used in pacemakers, implantable neural prosthetic devices typically have power requirements that exceed the capability of reasonably-sized implantable batteries. Therefore, transcutaneous magnetic coupling remains the method of choice for powering implanted neural prostheses. Using the same inductive link for transfer of power and bidirectional telemetry is an attractive solution to powering and communicating with implanted devices thus avoiding percutaneous plugs, wires, or conduits. The Class-E power oscillator has been identified as a highly-efficient transmitter circuit for use as a means of transferring power to an implant. Although the high-Q nature of this topology makes rapid modulation difficult, it is feasible to use synchronous frequency-shift-keyed (FSK) modulation of the Class-E circuit thereby combining an efficient power transmitter with a highspeed data link. Using this method, the transmitter can be modulated on a cycle-by-cycle basis with little to no additional power loss. Within an implanted device, demodulation of the FSK transmitted carrier can be accomplished using a novel demodulation circuit
引用
收藏
页码:3376 / 3379
页数:4
相关论文
共 7 条